Hameg HM 407A Скачать руководство пользователя страница 9

9

Subject to change without notice

probe can be compensated by a higher amplitude setting, the
probe should always be used. The series impedance of the
probe provides a certain amount of protection for the input of
the vertical amplifier. Because of their separate manufacture,
all attenuator probes are only partially compensated, therefore
accurate compensation must be performed on the oscilloscope
(see Probe compensation ).

Standard attenuator probes on the oscilloscope normally reduce
its bandwidth and increase the rise time. In all cases where the
oscilloscope bandwidth must be fully utilized (e.g. for pulses
with steep edges) we strongly advise using the probes

 HZ51

(x10)

 HZ52 

(x10 HF) and

 HZ54 

(x1 and x10). This can save the

purchase of an oscilloscope with larger bandwidth.

The probes mentioned have a HF-calibration in addition to low
frequency calibration adjustment. Thus a group delay correction
to the upper limit frequency of the oscilloscope is possible with
the aid of an 1MHz calibrator, e.g.

 HZ60

.

In fact the bandwidth and rise time of the oscilloscope are not
noticeably changed with these probe types and the waveform
reproduction fidelity can even be improved because the probe
can be matched to the oscilloscopes individual pulse response.

If a x10 or x100 attenuator probe is used, DC input
coupling must always be used at voltages above 400V.
With AC coupling of low frequency signals, the
attenuation is no longer independent of frequency,
pulses can show pulse tilts. Direct voltages are
suppressed but load the oscilloscope input coupling
capacitor concerned. Its voltage rating is max. 400 V
(DC + peak AC). DC input coupling is therefore of quite
special importance with a x100 attenuation probe which
usually has a voltage rating of max. 1200 V (DC + peak
AC). A capacitor of corresponding capacitance and
voltage rating may be connected in series with the
attenuator probe input for blocking DC voltage (e.g. for
hum voltage measurement).

With all attenuator probes, the maximum AC input voltage
must be derated with frequency usually above 20kHz. Therefore
the derating curve of the attenuator probe type concerned
must be taken into account. The selection of the ground point
on the test object is important when displaying small signal
voltages. It should always be as close as possible to the
measuring point. If this is not done, serious signal distortion
may result from spurious currents through the ground leads or
chassis parts. The ground leads on attenuator probes are also
particularly critical. They should be as short and thick as
possible. When the attenuator probe is connected to a BNC-
socket, a BNC-adapter, should be used. In this way ground and
matching problems are eliminated. Hum or interference
appearing in the measuring circuit (especially when a small
deflection coefficient is used) is possibly caused by multiple
grounding because equalizing currents can flow in the shielding
of the test cables (voltage drop between the protective
conductor connections, caused by external equipment
connected to the mains/line, e.g. signal generators with
interference protection capacitors).

Controls and readout

The following description assumes that the operating mode
“COMPONENT  TEST” is switched off. All important measuring
parameter settings are displayed in the screen readout when
the oscilloscope is on.

The LED indicators on the large front panel facilitate operation
and provide additional information. Electrical end positions of
controls are indicated by acoustic signal (beep).

All controls, except the power switch (POWER), the calibration
frequency pushbutton (CAL. 1kHz/1MHz), the FOCUS control
and the trace rotation control, are electronically set and
interrogated. Thus, all electronically set functions and their
current settings can be stored and also remotely controlled.
Some controls are only operative in the digital mode or have a
different function. Explanations pertaining to them are indicated
with the hint “storage mode only”.

The large front panel is, as is usual with Hameg oscilloscopes,
is marked with several fields.

The following controls and LED indicators are located on the
top, to the right of the screen, above the horizontal line:

(1)

POWER 

- Pushbutton and symbols for

 ON (I) 

and

 OFF (O)

.

After the oscilloscope is switched on, all LEDs are lit and
an automated instrument test is performed. During this
time the

 HAMEG 

logo and the software version are

displayed on the screen. After the internal test is completed
successfully, the overlay is switched off and the normal
operation mode is present. Then the last used settings
become activated and one LED indicates the ON condition.

It is possible to modify certain functions (SETUP) or to call
automatic calibration procedures (CALIBRATE).

 For details

relating to this see section “MENU”.

(2)

AUTOSET 

- Pushbutton

Briefly depressing this pushbutton results in an automatic
instrument setting automatically selecting Yt mode. The
instrument is set to the last used Yt mode setting (

CH I

,

CH II 

or

 DUAL

).

SEARCH (SEA) 

and

 DELAY 

(

DEL 

and

 DTR

) mode is

automatically switched off.

 Please note “AUTO SET”.

Automatic CURSOR supported voltage measurement

If 

CURSOR

 voltage measurement is present, the 

CURSOR

lines are automatically set to the positive and negative
peak value of the signal. The accuracy of this function
depends on the signal frequency and is also influenced by
the signal‘s pulse duty factor. If the signal height is
insufficient, the 

CURSOR

 lines do not change. In DUAL

mode the 

CURSOR

 lines are related to the signal which

is used for internal triggering.

STORAGE MODE ONLY

Additionally,

 AUTO SET

 automatically selects refresh

mode 

(RFR) 

when

 SINGLE (SGL) 

or

 ROLL (ROL) 

function

is in operation.

Automatic CURSOR supported measurement

In contrast to analog mode, 

AUTO SET

 also causes an

automatic 

CURSOR

 line setting if time or frequency

measurement has been selected and at least one signal
period is displayed. Neither the signal frequency nor the
pulse duty factor have an effect on the accuracy when
CURSOR voltage measurement is chosen.

(3)

RM - LED

The remote control mode can be switched on or off
(

”RM” 

LED dark) via the RS232 interface. On condition

Controls and readout

Содержание HM 407A

Страница 1: ...in analog mode 29 Auto Set 31 Save Recall 31 Component Tester analog mode 31 General 31 Using the Component Tester 32 Test Procedure 32 Test Pattern Displays 32 Testing Resistors 32 Testing Capacitors...

Страница 2: ...mity test made by HAMEG is based on the actual generic and product standards In cases where different limit values are applicable HAMEG applies the severer standard For emission the limits for residen...

Страница 3: ...100MHz 0 5div Normal DC 100MHz LED for trigger indication Slope positive or negative Sources CH I or II line ext CH I alternate CH II 0 8div Coupling AC 10Hz 100MHz DC 0 100MHz HF 50kHz 100MHz LF 0 1...

Страница 4: ...protective earth contact The protective action must not be negated by the use of an extension cord without a protective conductor Themains lineplugmustbeinsertedbeforeconnections are made to measuring...

Страница 5: ...is stated for possible queries this helps towards speeding up the processing of guarantee claims Maintenance Various important properties of the oscilloscope should be carefully checked at certain int...

Страница 6: ...he readout with the symbol for DC and the symbol for AC coupling Amplitude Measurements In general electrical engineering alternating voltage data normally refers to effective values rms root mean squ...

Страница 7: ...ore the measurement It can lie below or above the horizon tal central line according to whether positive and or negative deviations from the ground potential are to be measured Total value of input vo...

Страница 8: ...has little meaning For amplifiers with approximately constant group delay therefore good pulse transmission performance the following numerical relationship between rise time tr in ns and bandwidth B...

Страница 9: ...tective conductor connections caused by external equipment connected to the mains line e g signal generators with interference protection capacitors Controls and readout The following description assu...

Страница 10: ...a pre or post trigger value PT is not indicated by the readout Pressing and holding the STOR ON button switches over to the digital mode but without changing the channel operating mode CH I CH II DUAL...

Страница 11: ...and back to PT0 The values refer to the X axis graticule of the screen display 10 1div The following description assumes that the X magnifier x10 is inactive and the signal display starts on the left...

Страница 12: ...V In this case also several signal acquisition scans are required hence it is similar to Refresh operation The signal is averaged over the several acquisitions so that amplitude variations e g noise a...

Страница 13: ...r pretrigger and post trigger settings the signal acquisition is not complete when the trigger occurs and will only be terminated later After termination the RES LED extinguishes but the signal displa...

Страница 14: ...nob In ADD addition mode both Y POS I and Y POS II control knobs are active If the instrument is set to analog XY mode this control knob is inactive and the X POS knob must be used for a horizontal po...

Страница 15: ...trigger point setting voltage The trigger unit starts the time base when the edge of a trigger signal crosses the trigger point In most Yt modes the trigger point is displayed in the readout by the s...

Страница 16: ...nged All controls related to both channels are active if the inputs 31 and 35 are not set to GD 33 37 Whether alternated or chopped channel switching is present depends on the actual time base setting...

Страница 17: ...ing in DUAL mode Under these conditions both TRIG CH I and CH II LEDs are lit As alternate triggering requires alternate channel operation alternate channel switching is set automatically A change of...

Страница 18: ...of the trace can be set with the DEL POS control in the time base modes SEA SEARCH or DEL DELAY See SEA DEL ON OFF 29 STORAGE MODE ONLY The DEL POS control and the HO LED are disabled since in this m...

Страница 19: ...eviously active settings Automatic normal triggering 15 trigger LEVEL 17 trigger slope 15 and trigger coupling 26 will be stored The instrument will be automatically switched to normal triggering NM a...

Страница 20: ...as an earth ground symbol instead of the deflection coefficient and the AC or DC symbol The GD setting disables the input signal the AC DC 36 pushbutton and the VOLTS DIV 24 knob In automatic trigger...

Страница 21: ...In this mode two signals are normally applied causing an X and a Y deflection The deflection coefficient selected for each channel may be different thus as in DUAL mode the V cursor measurement requi...

Страница 22: ...ent the cursor moves fast 44 CAL Pushbutton and concentric socket A squarewave signal of 0 2Vpp 1 is available from the socket for probe adjustment purposes The signal frequency depends on the pushbut...

Страница 23: ...eady for use If the AUTO SET function was not used and only a spot appears CAUTION CRT phosphor can be damaged reduce the intensity immediately and check that the XY mode is not selected XY not displa...

Страница 24: ...displayed on the CRT screen should have the same value as during the 1kHz adjustment Probes other than those mentioned above normally have a larger tip diameter and may not fit into the calibrator out...

Страница 25: ...n the result The following must be noted here Becauseoftheperiodicnatureofthetrigonometricfunctions the calculation should be limited to angles 90 However here is the advantage of the method Due to ph...

Страница 26: ...g signal AC voltage are displayable in Yt mode In this mode the signal voltage deflects the beam in vertical direction Y while the time base generator moves the beam from the left to the right of the...

Страница 27: ...e previous slope setting for the undelayed time base trigger is stored and still active For further information please note Controls and readout The time base generator can be triggered by a rising or...

Страница 28: ...lit 26 The trigger point symbol is inactive in line mains trigger mode as there is no direct amplitude relationship between the trigger voltage and the signal voltage A voltage originating from mains...

Страница 29: ...sweep periods approx up to the ratio 10 1 Pulses or other signal waveforms appearing during this off period cannot trigger the time base Particularly with burst signals or aperiodic pulse trains of t...

Страница 30: ...tion coefficient must be increased with TIME DIV knob A larger deflection coefficient than in the SEARCH mode cannot be set Example The SEARCH setting selected in figure 2 is 5ms cm The display in DEL...

Страница 31: ...veral frequencies e g video signals are present AUTO SET sets the instrument automatically to the following operating conditions The input coupling is not affected while AC or DC is used but if GD was...

Страница 32: ...e will tend towards the horizontal axis and with low values the slope will move towards the vertical axis Values of resistance from 20 to 4 7k can be approximately evaluated The determination of actua...

Страница 33: ...n circuit tests are possible in many cases However they are not well defined This is caused by a shunt connection of real or complex impedances especially if they are of relatively low impedance at 50...

Страница 34: ...rizontal center graticule line by using the Y POS control Then the trigger point symbol may be set 2 division above the 0 Volt position using the LEVEL control If the deflection coefficient is set to...

Страница 35: ...commended to switch on the instrument for about 20 minutes prior to the commencement of any check Cathode Ray Tube Brightness and Focus Linearity Raster Distortion Normally the CRT of the instrument h...

Страница 36: ...regular intervals the deflection coefficients on all positions of the input attenuators and readjust them as necessary A compensated 2 1 series attenuator is also necessary and this must be matched t...

Страница 37: ...For precise determination set the peak of the first marker or cycle peak exactly behind the first vertical graticule line using the X POS control Deviation tendencies can be noted after some of the m...

Страница 38: ...ariable resistor 100k located on the CRT PCB is used for this adjustment procedure It may only be touched by a properlyinsulatingscrewdriver Caution Highvoltage Correct adjustment is achieved when the...

Страница 39: ...a high level reference potential is not permitted and endangers operator oscilloscope interface and peripheral devices In case of disregard of the safety warnings contained in this manual HAMEG refuse...

Страница 40: ...40 Subject to change without notice Front Panel HM407...

Отзывы: