Hameg HM 407A Скачать руководство пользователя страница 36

36

Subject to change without notice

wave signal of 10-100kHz is applied to the amplifier input.
When the

 Y-POS. 

control is then turned fully in both directions

from stop to stop with a display height of approximately 8div,
the upper and lower positions of the trace that are visible
should be approximately of the same height. Differences of up
to 1div are permissible (input coupling should be set to AC).

Checking the drift is relatively simple. 20minutes after switching
on the instrument, set the baseline exactly on the horizontal
center line of the graticule. The beam position must not
change by more than 0.5div during the following hour.

Calibration of the Vertical Amplifier

A square-wave voltage of 0.2Vpp ±1% is present at the output
socket of the calibrator (CAL.) If a direct connection is made
between the 0.2V output and the input of the vertical amplifier
(e.g. using a x1 probe), the displayed signal in the 50mV/div
position (variable control to CAL.) should be 4div high (DC input
coupling).

Maximum deviations of 0.12div (3%) are permissible. If a x10
probe (1%) is connected between the 0.2V output and Y input,
the deflection coefficient should be set to 5mV/div. Then the
maximum deviation is 0.16div.

With higher tolerances it should first be investigated whether
the cause lies, within the amplifier or in the amplitude of the
square-wave signal. On occasions it is possible that the probe
is faulty or incorrectly compensated. If necessary the measuring
amplifier can be calibrated with an accurately known DC
voltage (DC input coupling). The trace position should then
vary in accordance with the deflection coefficient set.

With variable control in the attenuator sector fully counter-
clockwise, the input sensitivity is decreased at least by the
factor 2.5 in each position. In the 50mV/div position, the
displayed calibrator signal height should vary from 4div to at
least 1.6div.

Transmission Performance
of the Vertical Amplifier

The transient response and the delay distortion correction can
only be checked with the aid of a square-wave generator with
a fast risetime (max. 5ns). The signal coaxial cable (e.g.

 HZ34

)

must be terminated at the vertical input of the oscilloscope
with a resistor equal to the characteristic impedance of the
cable (e.g. with HZ22). Checks should be made at 100Hz,
1kHz, 10kHz, 100kHz and 1MHz, the deflection coefficient
should be set at 5mV/div with DC input coupling. In so doing,
the square pulses must have a flat top without ramp-off,
spikes and glitches; no overshoot is permitted, especially at
1MHz and a display height of 4-5div. At the same time, the
leading top corner of the pulse must not be rounded. In
general, no great changes occur after the instrument has left
the factory, and it is left to the operators discretion whether
this test is undertaken or not. A suited generator for this test
is HZ60 from HAMEG.

Of course, the quality of the transmission performance is not
only dependent on the vertical amplifier. The input attenuators,
located in the front of the amplifier, are frequency-compensated
in each position. Even small capacitive changes can reduce the
transmission performance. Faults of this kind are as a rule most
easily detected with a square-wave signal with a low repetition
rate (e.g. 1kHz). If a suitable generator with max. output of
40Vpp is available, it is advisable to check at regular intervals the
deflection coefficients on all positions of the input attenuators
and readjust them as necessary. A compensated 2:1 series
attenuator is also necessary, and this must be matched to the

input impedance of the oscilloscope. This attenuator can be
made up locally. It is important that this attenuator is shielded.
For local manufacture, the electrical components required are
a 1M

 ±1% resistor and, in parallel with it, a trimmer 3-15pF in

parallel with approx. 10pF. One side of this parallel circuit is
connected directly to the input connector of CH I or CH II and
the other side is connected to the generator, if possible via a
low-capacitance coaxial cable. The series attenuator must be
matched to the input impedance of the oscilloscope in the 5mV/
div position (variable control to CAL., DC input coupling; square
tops exactly horizontal; no ramp-off is permitted). This is achieved
by adjusting the trimmer located in the 2:1 attenuator. The
shape of the square-wave should then be the same in each input
attenuator position.

Operating Modes:

CH.I/II, DUAL, ADD, CHOP.,

INVERT and X-Y Operation

In DUAL mode two traces must appear immediately. On
actuation of the Y-POS. controls, the trace positions should
have minimal effect on each other. Nevertheless, this cannot
be entirely avoided, even in fully serviceable instruments.
When one trace is shifted vertically across the entire screen,
the position of the other trace must not vary by more than
0.5mm.

A criterion in chopped operation is trace widening and
shadowing around and within the two traces in the upper or
lower region of the screen. Set time coefficient to 0.5ms/div,
set input coupling of both channels to GD and advance the

INTENS. 

control fully clockwise. Adjust

 FOCUS 

for a sharp

display. With the Y-POS. controls shift one of the traces to a
+2div, the other to a -2div vertical position from the horizontal
center line of the graticule.

Do not try to synchronize (with the time variable
control) the chop frequency (0.5MHz)! Check for negli-
gible trace widening and periodic shadowing when
switching between 0.5ms/div and 0.2ms/div.

It is important to note that in the I+II add mode or the I-II
difference mode the vertical position of the trace can be adjusted
by using both the

 Channel I 

and

 Channel II Y-POS.

 controls.

In X-Y Operation, the sensitivity in both deflection directions
will be the same. When the signal from the built-in square-
wave generator is applied to the input of Channel II, then, as
with Channel I in the vertical direction, there must be a
horizontal deflection of 4div when the deflection coefficient is
set to 50mV/div position. The check of the mono channel
display is unnecessary; it is contained indirectly in the tests
above stated.

Triggering Checks

The internal trigger threshold is important as it determines the
display height from which a signal will be stably displayed. It
should be approx. 0.3-0.5div for the instrument. An increased
trigger sensitivity creates the risk of response to the noise
level in the trigger circuit. This can produce double-triggering
with two out-of-phase traces.

Alteration of the trigger threshold is not required. Checks can
be made with any sine-wave voltage between 50Hz and
1MHz. The instrument should be in automatic peak (value)
triggering (NM LED dark) and the

 LEVEL 

knob in electrical

midrange position. It should be ascertained whether the same
trigger sensitivity is also present with Normal Triggering (

NM

LED 

lights). In this trigger mode, LEVEL adjustment is absolutely

necessary.

Test Instructions

Содержание HM 407A

Страница 1: ...in analog mode 29 Auto Set 31 Save Recall 31 Component Tester analog mode 31 General 31 Using the Component Tester 32 Test Procedure 32 Test Pattern Displays 32 Testing Resistors 32 Testing Capacitors...

Страница 2: ...mity test made by HAMEG is based on the actual generic and product standards In cases where different limit values are applicable HAMEG applies the severer standard For emission the limits for residen...

Страница 3: ...100MHz 0 5div Normal DC 100MHz LED for trigger indication Slope positive or negative Sources CH I or II line ext CH I alternate CH II 0 8div Coupling AC 10Hz 100MHz DC 0 100MHz HF 50kHz 100MHz LF 0 1...

Страница 4: ...protective earth contact The protective action must not be negated by the use of an extension cord without a protective conductor Themains lineplugmustbeinsertedbeforeconnections are made to measuring...

Страница 5: ...is stated for possible queries this helps towards speeding up the processing of guarantee claims Maintenance Various important properties of the oscilloscope should be carefully checked at certain int...

Страница 6: ...he readout with the symbol for DC and the symbol for AC coupling Amplitude Measurements In general electrical engineering alternating voltage data normally refers to effective values rms root mean squ...

Страница 7: ...ore the measurement It can lie below or above the horizon tal central line according to whether positive and or negative deviations from the ground potential are to be measured Total value of input vo...

Страница 8: ...has little meaning For amplifiers with approximately constant group delay therefore good pulse transmission performance the following numerical relationship between rise time tr in ns and bandwidth B...

Страница 9: ...tective conductor connections caused by external equipment connected to the mains line e g signal generators with interference protection capacitors Controls and readout The following description assu...

Страница 10: ...a pre or post trigger value PT is not indicated by the readout Pressing and holding the STOR ON button switches over to the digital mode but without changing the channel operating mode CH I CH II DUAL...

Страница 11: ...and back to PT0 The values refer to the X axis graticule of the screen display 10 1div The following description assumes that the X magnifier x10 is inactive and the signal display starts on the left...

Страница 12: ...V In this case also several signal acquisition scans are required hence it is similar to Refresh operation The signal is averaged over the several acquisitions so that amplitude variations e g noise a...

Страница 13: ...r pretrigger and post trigger settings the signal acquisition is not complete when the trigger occurs and will only be terminated later After termination the RES LED extinguishes but the signal displa...

Страница 14: ...nob In ADD addition mode both Y POS I and Y POS II control knobs are active If the instrument is set to analog XY mode this control knob is inactive and the X POS knob must be used for a horizontal po...

Страница 15: ...trigger point setting voltage The trigger unit starts the time base when the edge of a trigger signal crosses the trigger point In most Yt modes the trigger point is displayed in the readout by the s...

Страница 16: ...nged All controls related to both channels are active if the inputs 31 and 35 are not set to GD 33 37 Whether alternated or chopped channel switching is present depends on the actual time base setting...

Страница 17: ...ing in DUAL mode Under these conditions both TRIG CH I and CH II LEDs are lit As alternate triggering requires alternate channel operation alternate channel switching is set automatically A change of...

Страница 18: ...of the trace can be set with the DEL POS control in the time base modes SEA SEARCH or DEL DELAY See SEA DEL ON OFF 29 STORAGE MODE ONLY The DEL POS control and the HO LED are disabled since in this m...

Страница 19: ...eviously active settings Automatic normal triggering 15 trigger LEVEL 17 trigger slope 15 and trigger coupling 26 will be stored The instrument will be automatically switched to normal triggering NM a...

Страница 20: ...as an earth ground symbol instead of the deflection coefficient and the AC or DC symbol The GD setting disables the input signal the AC DC 36 pushbutton and the VOLTS DIV 24 knob In automatic trigger...

Страница 21: ...In this mode two signals are normally applied causing an X and a Y deflection The deflection coefficient selected for each channel may be different thus as in DUAL mode the V cursor measurement requi...

Страница 22: ...ent the cursor moves fast 44 CAL Pushbutton and concentric socket A squarewave signal of 0 2Vpp 1 is available from the socket for probe adjustment purposes The signal frequency depends on the pushbut...

Страница 23: ...eady for use If the AUTO SET function was not used and only a spot appears CAUTION CRT phosphor can be damaged reduce the intensity immediately and check that the XY mode is not selected XY not displa...

Страница 24: ...displayed on the CRT screen should have the same value as during the 1kHz adjustment Probes other than those mentioned above normally have a larger tip diameter and may not fit into the calibrator out...

Страница 25: ...n the result The following must be noted here Becauseoftheperiodicnatureofthetrigonometricfunctions the calculation should be limited to angles 90 However here is the advantage of the method Due to ph...

Страница 26: ...g signal AC voltage are displayable in Yt mode In this mode the signal voltage deflects the beam in vertical direction Y while the time base generator moves the beam from the left to the right of the...

Страница 27: ...e previous slope setting for the undelayed time base trigger is stored and still active For further information please note Controls and readout The time base generator can be triggered by a rising or...

Страница 28: ...lit 26 The trigger point symbol is inactive in line mains trigger mode as there is no direct amplitude relationship between the trigger voltage and the signal voltage A voltage originating from mains...

Страница 29: ...sweep periods approx up to the ratio 10 1 Pulses or other signal waveforms appearing during this off period cannot trigger the time base Particularly with burst signals or aperiodic pulse trains of t...

Страница 30: ...tion coefficient must be increased with TIME DIV knob A larger deflection coefficient than in the SEARCH mode cannot be set Example The SEARCH setting selected in figure 2 is 5ms cm The display in DEL...

Страница 31: ...veral frequencies e g video signals are present AUTO SET sets the instrument automatically to the following operating conditions The input coupling is not affected while AC or DC is used but if GD was...

Страница 32: ...e will tend towards the horizontal axis and with low values the slope will move towards the vertical axis Values of resistance from 20 to 4 7k can be approximately evaluated The determination of actua...

Страница 33: ...n circuit tests are possible in many cases However they are not well defined This is caused by a shunt connection of real or complex impedances especially if they are of relatively low impedance at 50...

Страница 34: ...rizontal center graticule line by using the Y POS control Then the trigger point symbol may be set 2 division above the 0 Volt position using the LEVEL control If the deflection coefficient is set to...

Страница 35: ...commended to switch on the instrument for about 20 minutes prior to the commencement of any check Cathode Ray Tube Brightness and Focus Linearity Raster Distortion Normally the CRT of the instrument h...

Страница 36: ...regular intervals the deflection coefficients on all positions of the input attenuators and readjust them as necessary A compensated 2 1 series attenuator is also necessary and this must be matched t...

Страница 37: ...For precise determination set the peak of the first marker or cycle peak exactly behind the first vertical graticule line using the X POS control Deviation tendencies can be noted after some of the m...

Страница 38: ...ariable resistor 100k located on the CRT PCB is used for this adjustment procedure It may only be touched by a properlyinsulatingscrewdriver Caution Highvoltage Correct adjustment is achieved when the...

Страница 39: ...a high level reference potential is not permitted and endangers operator oscilloscope interface and peripheral devices In case of disregard of the safety warnings contained in this manual HAMEG refuse...

Страница 40: ...40 Subject to change without notice Front Panel HM407...

Отзывы: