Hameg HM 407A Скачать руководство пользователя страница 25

25

Subject to change without notice

Antiphase input voltages:

    No invert function active = difference.
    Channel II invert function active = sum.

In the

 ADD 

mode the vertical display position is dependent

upon the

 Y-POS. 

setting of both channels. The same Y

deflection coefficient is normally used for both channels with
algebraic addition.

Please note that the

 Y-POS. 

settings are also added but are not

affected by the

 INV 

setting.

Differential measurement techniques allow direct measure-
ment of the voltage drop across floating components (both
ends above ground). Two identical probes should be used for
both vertical inputs. In order to avoid ground loops, use a
separate ground connection and do not use the probe ground
leads or cable shields.

X-Y Operation

The important control  for this mode is the pushbutton labeled

DUAL 

and

 XY (22)

. The following description refers to the

 XY

analog mode.

In

 XY 

mode the time base is deactivated. The signal applied to

the input of channel I - front panel marking

 HOR. INP. (X) 

-

causes the X deflection. The input related controls (

AC/DC

,

GD

 pushbutton and the

 VOLTS/DIV 

knob) consequently

affect the X deflection. For X position alteration, the X-POS.
control knob must be used, as the

 Y-POS. I 

control is

automatically inactivated. The input deflection coefficient ranges
are the same for both channels, because the

 X-MAG x10

function is inactive in XY mode.

The bandwidth of the X amplifier, is lower than the Y amplifier
and the phase angle which increases with higher frequencies,
must be taken into account (please note data sheet).

  

The inversion of the X-input signal is not possible.

Lissajous figures can be displayed in the X-Y mode for certain
measuring tasks:

• Comparing two signals of different frequency or bringing

one frequency up to the frequency of the other signal. This
also applies for whole number multiples or fractions of the
one signal frequency.

• Phase comparison between two signals of the same

frequency.

Phase comparison with Lissajous figures

The following diagrams show two sine signals of the same
frequency and amplitude with different phase angles.

Calculation of the phase angle or the phase shift between the
X and Y input voltages (after measuring the distances a and b
on the screen) is quite simple with the following formula, and
a pocket calculator with trigonometric functions. Apart from
the reading accuracy, the signal height has no influence on the
result.

The following must be noted here:
• Because of the periodic nature of the trigonometric functions,

the calculation should be limited to angles 

90°. However

here is the advantage of the method.

• Due to phase shift, do not use a too high test frequency.
• It cannot be seen as a matter of course from the screen

display if the test voltage leads or lags the reference
voltage. A CR network before the test voltage input of the
oscilloscope can help here. The 1 M

 input resistance can

equally serve as R here, so that only a suitable capacitor C
needs to be connected in series. If the aperture width of the
ellipse is increased (compared with C short-circuited), then
the test voltage leads the reference voltage and vice versa.
This applies only in the region up to 90° phase shift.
Therefore C should be sufficiently large and produce only a
relatively small just observable phase shift.

Should both input voltages be missing or fail in the X-Y
mode, a very bright light dot is displayed on the screen.
This dot can burn into the phosphor at a too high
brightness setting (INTENS. knob) which causes either
a lasting loss of brightness, or in the extreme case,
complete destruction of the phosphor at this point.

Phase difference measurement
in DUAL mode (Yt)

Phase differences between two input signals of the same
frequency and shape can be measured very simply on the
screen in Dual mode. The time base should be triggered by the
reference signal (phase position 0).  The other signal can then
have a leading or lagging phase angle. In alternate triggering
condition, phase difference measurement is not possible. For
greatest accuracy adjust the time base for slightly over one
period and approximately the same height of both signals on
the screen. The Y deflection coefficients, the time base
coefficients and the trigger level setting can be used for this
adjustment, without influence on the result. Both base lines
are set onto the horizontal graticule center line using the Y-
POS. knobs before the measurement. With sinusoidal signals,
observe the zero (crossover point) transitions; the sine peaks
are less accurate. If a sine signal is noticeably distorted by even
harmonics, or if a DC voltage is present, AC coupling is
recommended for both channels. If it is a question of pulses
of the same shape, read off at steep edges.

It must be noted that the phase difference cannot be determined
if alternate triggering is selected.

Phase difference measurement in DUAL mode

t = horizontal spacing of the zero transitions in div
T = horizontal spacing for one period in div

Operating modes of the vertical amplifiers in Yt mode

Содержание HM 407A

Страница 1: ...in analog mode 29 Auto Set 31 Save Recall 31 Component Tester analog mode 31 General 31 Using the Component Tester 32 Test Procedure 32 Test Pattern Displays 32 Testing Resistors 32 Testing Capacitors...

Страница 2: ...mity test made by HAMEG is based on the actual generic and product standards In cases where different limit values are applicable HAMEG applies the severer standard For emission the limits for residen...

Страница 3: ...100MHz 0 5div Normal DC 100MHz LED for trigger indication Slope positive or negative Sources CH I or II line ext CH I alternate CH II 0 8div Coupling AC 10Hz 100MHz DC 0 100MHz HF 50kHz 100MHz LF 0 1...

Страница 4: ...protective earth contact The protective action must not be negated by the use of an extension cord without a protective conductor Themains lineplugmustbeinsertedbeforeconnections are made to measuring...

Страница 5: ...is stated for possible queries this helps towards speeding up the processing of guarantee claims Maintenance Various important properties of the oscilloscope should be carefully checked at certain int...

Страница 6: ...he readout with the symbol for DC and the symbol for AC coupling Amplitude Measurements In general electrical engineering alternating voltage data normally refers to effective values rms root mean squ...

Страница 7: ...ore the measurement It can lie below or above the horizon tal central line according to whether positive and or negative deviations from the ground potential are to be measured Total value of input vo...

Страница 8: ...has little meaning For amplifiers with approximately constant group delay therefore good pulse transmission performance the following numerical relationship between rise time tr in ns and bandwidth B...

Страница 9: ...tective conductor connections caused by external equipment connected to the mains line e g signal generators with interference protection capacitors Controls and readout The following description assu...

Страница 10: ...a pre or post trigger value PT is not indicated by the readout Pressing and holding the STOR ON button switches over to the digital mode but without changing the channel operating mode CH I CH II DUAL...

Страница 11: ...and back to PT0 The values refer to the X axis graticule of the screen display 10 1div The following description assumes that the X magnifier x10 is inactive and the signal display starts on the left...

Страница 12: ...V In this case also several signal acquisition scans are required hence it is similar to Refresh operation The signal is averaged over the several acquisitions so that amplitude variations e g noise a...

Страница 13: ...r pretrigger and post trigger settings the signal acquisition is not complete when the trigger occurs and will only be terminated later After termination the RES LED extinguishes but the signal displa...

Страница 14: ...nob In ADD addition mode both Y POS I and Y POS II control knobs are active If the instrument is set to analog XY mode this control knob is inactive and the X POS knob must be used for a horizontal po...

Страница 15: ...trigger point setting voltage The trigger unit starts the time base when the edge of a trigger signal crosses the trigger point In most Yt modes the trigger point is displayed in the readout by the s...

Страница 16: ...nged All controls related to both channels are active if the inputs 31 and 35 are not set to GD 33 37 Whether alternated or chopped channel switching is present depends on the actual time base setting...

Страница 17: ...ing in DUAL mode Under these conditions both TRIG CH I and CH II LEDs are lit As alternate triggering requires alternate channel operation alternate channel switching is set automatically A change of...

Страница 18: ...of the trace can be set with the DEL POS control in the time base modes SEA SEARCH or DEL DELAY See SEA DEL ON OFF 29 STORAGE MODE ONLY The DEL POS control and the HO LED are disabled since in this m...

Страница 19: ...eviously active settings Automatic normal triggering 15 trigger LEVEL 17 trigger slope 15 and trigger coupling 26 will be stored The instrument will be automatically switched to normal triggering NM a...

Страница 20: ...as an earth ground symbol instead of the deflection coefficient and the AC or DC symbol The GD setting disables the input signal the AC DC 36 pushbutton and the VOLTS DIV 24 knob In automatic trigger...

Страница 21: ...In this mode two signals are normally applied causing an X and a Y deflection The deflection coefficient selected for each channel may be different thus as in DUAL mode the V cursor measurement requi...

Страница 22: ...ent the cursor moves fast 44 CAL Pushbutton and concentric socket A squarewave signal of 0 2Vpp 1 is available from the socket for probe adjustment purposes The signal frequency depends on the pushbut...

Страница 23: ...eady for use If the AUTO SET function was not used and only a spot appears CAUTION CRT phosphor can be damaged reduce the intensity immediately and check that the XY mode is not selected XY not displa...

Страница 24: ...displayed on the CRT screen should have the same value as during the 1kHz adjustment Probes other than those mentioned above normally have a larger tip diameter and may not fit into the calibrator out...

Страница 25: ...n the result The following must be noted here Becauseoftheperiodicnatureofthetrigonometricfunctions the calculation should be limited to angles 90 However here is the advantage of the method Due to ph...

Страница 26: ...g signal AC voltage are displayable in Yt mode In this mode the signal voltage deflects the beam in vertical direction Y while the time base generator moves the beam from the left to the right of the...

Страница 27: ...e previous slope setting for the undelayed time base trigger is stored and still active For further information please note Controls and readout The time base generator can be triggered by a rising or...

Страница 28: ...lit 26 The trigger point symbol is inactive in line mains trigger mode as there is no direct amplitude relationship between the trigger voltage and the signal voltage A voltage originating from mains...

Страница 29: ...sweep periods approx up to the ratio 10 1 Pulses or other signal waveforms appearing during this off period cannot trigger the time base Particularly with burst signals or aperiodic pulse trains of t...

Страница 30: ...tion coefficient must be increased with TIME DIV knob A larger deflection coefficient than in the SEARCH mode cannot be set Example The SEARCH setting selected in figure 2 is 5ms cm The display in DEL...

Страница 31: ...veral frequencies e g video signals are present AUTO SET sets the instrument automatically to the following operating conditions The input coupling is not affected while AC or DC is used but if GD was...

Страница 32: ...e will tend towards the horizontal axis and with low values the slope will move towards the vertical axis Values of resistance from 20 to 4 7k can be approximately evaluated The determination of actua...

Страница 33: ...n circuit tests are possible in many cases However they are not well defined This is caused by a shunt connection of real or complex impedances especially if they are of relatively low impedance at 50...

Страница 34: ...rizontal center graticule line by using the Y POS control Then the trigger point symbol may be set 2 division above the 0 Volt position using the LEVEL control If the deflection coefficient is set to...

Страница 35: ...commended to switch on the instrument for about 20 minutes prior to the commencement of any check Cathode Ray Tube Brightness and Focus Linearity Raster Distortion Normally the CRT of the instrument h...

Страница 36: ...regular intervals the deflection coefficients on all positions of the input attenuators and readjust them as necessary A compensated 2 1 series attenuator is also necessary and this must be matched t...

Страница 37: ...For precise determination set the peak of the first marker or cycle peak exactly behind the first vertical graticule line using the X POS control Deviation tendencies can be noted after some of the m...

Страница 38: ...ariable resistor 100k located on the CRT PCB is used for this adjustment procedure It may only be touched by a properlyinsulatingscrewdriver Caution Highvoltage Correct adjustment is achieved when the...

Страница 39: ...a high level reference potential is not permitted and endangers operator oscilloscope interface and peripheral devices In case of disregard of the safety warnings contained in this manual HAMEG refuse...

Страница 40: ...40 Subject to change without notice Front Panel HM407...

Отзывы: