20
7.2. Fault Indicator Block Diagram
Figure 7:
Fault Indicator Block Diagram
See below if any problems are encountered in operating the TLS-EQ-9.
Condition:
Controller Fault and/or Lamp House Fault LED(s) are ON.
Action:
Always begin operation of the source by verifying the interlocks.
Confirm that the external interlock contact is closed (or that the black jumper plug is fully inserted
into the back of the Remote Control Box)
To reset or clear a Fault condition, actuate the “Operate” switch from the ON position to the OFF
position. If a fault was generated while the “Operate” switch was in the OFF position, first actuate
the switch to the ON position, then to the OFF position. The unit will not turn on if a fault condition
exists.
If
all the interlocks are OK and either the lamp or controller interlock faults will not clear, please
contact Energetiq
.
Condition:
Lamp fails to ignite after several tries.
Action
:
Contact Energetiq.
Lamp Fault
Asserted
(pin 3 on remote)
Diagnosti
Safet
Bulb Interlock
Open
Lamp Head
Overtemp > 85C
Control Cable Not
Connected
Plasma
Extinguished
(after ignition)
Controller
Fault Asserted
(pin 4 on remote)
Laser Overtemp
(>60C)
Controller PCB
Overtemp (>82C)
Laser Current
Regulation Error
External
Interlock Open
Laser Temperature
Regulation Error
Signal Available on User Interface
Internal Logic That Generates Signal
Lamp On
De-Asserted
(pin 1 on remote)
Laser On
De-Asserted 150 sec
after Operate
Command
(pin 2 on remote)
Fail to Ignite