Chapter 2. System Description
EQ-99X Operation & Maintenance Manual
12
of
45
A. Status Indicator LEDs
LED Label
Meaning (when lit)
POWER ON
DC power is connected to the power supply controller
LAMP ON
UV Light is on
LASER ON
Laser power is ON and laser light is being delivered to the lamp head
CONTROLLER FAULT
One of the following has occurred in the power supply controller:
EQ-99X Power Supply Controller Components
Figure 4.
Status Indicator LED Functions
Table 2.
1. External interlock open
2. Controller internal temperature too high
3. Laser power not reaching setpoint
4. Laser temperature fault
5. Failure in laser delivery fiber