C9300-124 Ver.1.4
11
8. OPERATION
8-1 PRECAUTIONS
(1) Ambient temperature
This camera is cooled using a peltier element. In a peltier element, one side is cooled
and the other side heated by the flow of electric current. A CCD chip is installed on the
cooled side, and the heated side undergoes forced air-cooling. The recommended
ambient temperature is 20 °C.
Note
•
The temperature rises above 20 °C and operation becomes unstable.
8-2 PREPARATION FOR IMAGING
(1) Connect the equipment as shown in Fig.7-1 before starting operating of the
camera.
(2) Switch the camera's power switch to ON.
After turning the power switch on, wait over 30 minutes for aging; this completes
preparation for imaging.
8-3 IMAGING
The camera works with following setting after power is switched ON.
CCD output mode
Single output
Scan mode
Normal readout
Exposure time setting Normal
setting
Contrast enhancement gain
0
Digital data output bit count
12
The camera mode and parameters can be changed by command from the host computer,
which transmitted via the CameraLink interface.
8-4 END OF IMAGING
(1) End imaging or transmission of image data with the control software when imaging
is finished.
(2) Turn off the power to the camera and peripheral equipment.