C9300-124 Ver.1.4
9
The camera employs Hirose model HR10A-10R-12PB connector.
Number Signal
Pin
connection
1 GND
(Camera)
2
12 V (Camera)
3 NC
4 NC
5 NC
6 NC
7 NC
8 NC
9 NC
10 GND
(Peltier)
11
12 V (Peltier)
12 NC
HR10A-10R-12PB
③
Camera mount
Jig for fixing the camera. Used when mounting the camera on a tripod or other fixture.
Note
•
Be careful not to allow the fitting screw to enter more than 5 mm
from the surface of the mount. Screwing this in excessively may
impair normal operation.
④
CameraLink interface connector [DIGITAL OUT]
Connector used for connecting the camera to the CameraLink interface board.
⑤
Timing I/O connector [TIMING I/O]
The connector is used as the external trigger input (1
st
-pin, 2
nd
-pin).
•
Do not connect anything to pins from 3 to 6 because they are
reserved for options.
⑥
Mode switching switch
This switch is for the camera maintenance.
•
All positions are set to “OFF”. Do not change this.