(2) Parameter setting commands
(SHutter Time)
Parameter
n
(
n
: Horizontal line count)
・
Normal readout : Single output 1 to 6698
Dual output
1 to 12 285
SHT
Function
Sets the exposure time in units of 1H when the exposure trigger
timing is controlled internally and the electronic shutter is selected
as the method for setting the exposure time.
(External Shutter Time)
Parameter
n
(
n
: Horizontal line count)
・
Normal readout : Single output 1 to 6698
Dual output
1 to 12 285
EST
Function
Sets the exposure time in units of 1H when the exposure trigger
timing is controlled externally and the internal settings for exposure
time are used.
(Acquire Exposure Time)
Parameter
ss.xxxxxx
ss
: value of seconds,
xxxxxx
: value of microseconds
xx s
s
: seconds
xx ms
ms
: milliseconds
xx us
us
: microseconds
Function
Sets the exposure time by using the time when Free running mode
or the internal setting of External control mode.
The exposure time of 199 µs or less and 1 s or more cannot be set.
AET 0.123
Sets the exposure time at 123 ms.
AET
Example
AET 123ms
Sets the exposure time at 123 ms.
(Active Trigger Polarity)
Parameter
N
(Negative)
/
P
(Positive)
Function
Selects either Negative or Positive for the external trigger pulse
polarity.
ATP N
Sets the external trigger pulse polarity to Negative.
ATP
Example
ATP P
Sets the external trigger pulse polarity to Positive.
(Scan V-Offset)
Parameter
n
(0
≦
n
≦
2664,
n
=multiples of 8)
SVO
Function
Set the vertical start address for the sub-array readout. When the n
value is set to a number that is not a multiple of 8, the command
elicits an error (E3) in response.
(Scan V-Width)
Parameter
m
(8
≦
m
≦
2672,
m
=multiples of 8)
Function
Set the vertical line width for the sub-array readout. When the m
value is set to a number that is not a multiple of 8, the command
elicits an error (E3) in response.
SVW
Example
Offset 256 (start address 256)
Width 512
SVO 256
SVW 512
SMD A
512
256
(External trigger SourCe)
Parameter
M
(Multi Timing I/O)
/I
(I/F)
ESC
Function
Selects the input connector for external control pulse.
Содержание C9300-124
Страница 42: ...C9300 124 Ver 1 4 41 15 DIMENSIONAL OUTLINES 15 1 CAMERA Unit mm Depth 8 F mount Depth 8 ...
Страница 45: ...C9300 124 Ver 1 4 44 Technical sheet 1 C9300 124 Dimensional outline with Camera mount ...
Страница 46: ...C9300 124 Ver 1 4 45 Technical sheet 2 C9300 124 Dimensional outline without Camera mount ...