qCMOS Camera C15550-20UP / C15550-20UP01 Instruction manual_Ver.1.3
19
(5) Handling of the circulating water cooler
Handle the circulating water cooler and the cooling water according to an instruction manual of
the circulating water cooler.
Proper performance may not be achievable if a non-recommended circulating water cooler is used.
(6) Start water-cooling and water-cooling in operation
- Confirm the water is flowing before starting the camera cooling and that the camera does cool.
- Keep 0.45 L/min flow rate for water circulation.
- Do not stop the circulating water cooler while the camera is working.
(7) Cooling water hose
The hose has a blue hose (Internal diameter: 4 mm / External diameter: 6 mm) and a gray hose
(Internal diameter: 8 mm / External diameter: 13.5 mm). (Figure 8-1)
If the hose size on circulating water cooler is the same as blue hose, remove gray hose from the
joint part. The gray hose can be removed when blue hose is pulled with pushing the button of the
joint on gray hose. (Figure 8-2)
Figure 8-1
Figure 8-2
(8) Connection of the cooling water hose
•
Follow the instruction in Section 8-2 “CONNECTION OF WATER COOLING
HOSES” and Section 8-3 “DISCONNECTION OF WATER COOLING
HOSES” to connect / disconnect the hose.
- Stop water circulation when connecting / disconnecting the hose, and turn off the power of the
camera and the circulating water cooler.
- Confirm that cooling water stops.
- Prepare water absorption sheet (such as Waste, Towel or so) and catch pan in order to avoid
water drop or water splash.
(9) Deterioration of the cooling water hose
Replace the water hose with a new one whenever it cannot keep 0.45 L/min flow rate for water
circulation due to the hose deterioration.