C11440-22C Instruction manual_Ver.1.4
21
(3) Protection circuit
A double protection circuit protects this camera’s thermoelectric cooling device. If the
heat dissipater becomes abnormally hot, the protection circuit sets off a buzzer alarm
from the camera while simultaneously cutting the current supply to Peltier element.
When the protection circuit operates like this, immediately turn off the power switch.
Then locate and remove the cause of the overheating and restart.
9-2 PREPARATION FOR IMAGING
Use the following procedure when starting operating of the camera.
•
When you connect cables, turn off the power supply of the camera and the peripheral devices.
•
Please change cooling method (cooling mode) if needed. The present cooling mode set-up of
this camera can be checked using phxver.exe.
9-2-1
WHEN USING AIR-COOLING
(1) Connect the equipment as shown in Figure 7-1 before operating of the camera.
(2) Turn on the power to the camera.
(3) Check that the cooling fan is turning properly and that air is circulating.
(4) Switch to the cooling mode from phxver.exe.
Note
•
The cooling temperature becomes stable about 10 minutes after cooling begins.
9-2-2
WHEN USING WATER-COOLING
(1) Connect the equipment as shown in Figure 7-1 before operating of the camera.
(2) Turn on the power to the camera.
(3) Turn on the power to the circulating water cooler.
(4) Check that cooling water always circulates suitably.
Note
•
The cooling temperature becomes stable about 10 minutes after cooling begins.
(5) Turn on the cooling switch of the camera from application software.
Note
•
Please refer to the manual of application software for ON/OFF of the cooling
switch of a camera.
9-3 END OF IMAGING
Carry out the procedure below when imaging is finished.
(1) End the imaging or transmission of image data with the application software.
(2) Turn off the power to the circulating water cooler. (When set to water-cooled)
(3) Turn off the power to the camera and peripheral device.
Содержание C11440-22C
Страница 2: ...Blank page...
Страница 49: ...C11440 22C Instruction manual_Ver 1 4 47 15 DIMENSIONAL OUTLINES Unit mm...