【Saving the data】 

To save the data, stop the measurement and then select [Save] from the [File] menu (see Figure H) 
or click the 

  button on the toolbar. Then, in the dialog box that appears, specify a file name for 

saving the data. The data is saved in text format (*.txt). 

 

 

Figure H: Saving the data 

 
 

The sample software stores data acquired during measurement (from the start to stop of 
measurement) into the memory. That data is retained in the memory until the next measurement 
begins.  
Data saved with the [Save] command or button are the operating condition settings and number 
of input counts per gate time. 
(Example) If measurement is performed for 1 hour with a gate time of 10 ms, then a total of 
360,000 data items (1 hour / 10 ms = 360,000) will be acquired. 
The maximum number of data the memory can hold is 3,600,000 (this is equivalent to a 1-hour 
measurement with 1 ms gate time). If this value is exceeded, then new data will overwrite the old 
data starting from the oldest one. 

 
 

【Others: Graph manipulation】 

This sample software includes display tools such as zoom-in (

) of a selected area, auto scaling 

(

), vertical scale setting option (

), and cursor display (

). 

 

 

Figure I: Display tools 

 

【Quitting the software】 

To quit the software, select [Exit] from the [File] menu. 
 

Содержание C10507 Series

Страница 1: ...K29 B60467e MPPC Module C10507 series C10751 series INSTRUCTION MANUAL Doc Version 0 3 Hamamatsu Photonics K K 1126 1 Ichino cho Higashi ku Hamamatsu City 435 8558 Japan Telephone 053 434 3311...

Страница 2: ...n using this product In particular when this product is to be used in equipment or systems which might cause personal injury fatal accident or damage to property if handled improperly be sure to imple...

Страница 3: ...uch internal parts of this product Do not apply vibrations or shocks to this product Excessive shocks or vibrations may damage the internal components or adversely affect their adjustments causing fir...

Страница 4: ...imit on number of connectable modules 9 Setup 10 Installing the software 10 Installing the device driver 13 Uninstalling the software 15 Software operation 17 Starting the software 17 Measurement Star...

Страница 5: ...rwise the software installation may fail See the description in Setup for how to install the software Power to this product is supplied from the USB port on the PC Due to the USB specifications the ma...

Страница 6: ...requires high precision voltage control because the MPPC is used in Geiger mode where the gain greatly depends on changes in bias voltage and also because the usable bias voltage range is narrower tha...

Страница 7: ...e PC also makes various settings and monitors the counter value The counter value can also be obtained by connecting the comparator output to a frequency counter etc Connecting the analog signal to an...

Страница 8: ...om a minimum of 1 ms to a maximum of 100 ms Threshold level Setting the threshold level to one half 0 5 p e the height of the 1 p e pulse and then using a frequency counter to count the number of puls...

Страница 9: ...MPPC module to the PC before installing the software in the PC Otherwise the software might not be installed correctly See the section Installing the software 1 Connect the USB A plug of the USB cable...

Страница 10: ...verified on the following systems Operation on other systems is not guaranteed Microsoft Windows 2000 Professional SP4 Microsoft Windows XP Professional SP2 To operate the sample software you must us...

Страница 11: ...river manually before installing the software Installing the software After Windows has started insert the supplied CD ROM into the DVD CD ROM drive on the PC When the auto run for the CD is enabled t...

Страница 12: ...Complete default in most cases and then click Next The window for copying files appears It shows a list of components to be installed Click Next Make sure that the MPPC module is not connected to the...

Страница 13: ...12 This screen appears when the installation is complete Click Finish to end the installation wizard...

Страница 14: ...eat the same procedure to install the device driver Connect the MPPC module to the PC The MPPC will be detected as new hardware and the Found New Hardware Wizard will appear to guide you through the d...

Страница 15: ...compatible hardware check box is checked default select MPPC USB Module in the Model list and click Next to continue The device driver installation starts This screen appears when the installation is...

Страница 16: ...e Start menu select All Programs Hamamatsu MPPC Module Change or Remove When this screen appears select Remove then click Next When this confirmation message appears click OK If the MPPC is connected...

Страница 17: ...s screen appears when uninstallation is complete Click Finish After uninstallation this screen appears asking if you want to restart the PC Select Yes I want to restart my computer now and then click...

Страница 18: ...lecting the MPPC module to be used When the Option dialog box appears select the operating conditions from the Gate Time ms and Threshold Level drop down lists The gate time can be selected from 7 dif...

Страница 19: ...and stop To start measurement select Start from the Monitor menu as shown in Figure D or click the button on the toolbar To stop measurement select Stop from the Monitor menu see Figure E or click the...

Страница 20: ...its Shows the current option settings and temperature information The graph is a time series display and is updated every 100 ms Measurement data is thinned out and plotted on the graph Changing the o...

Страница 21: ...n are the operating condition settings and number of input counts per gate time Example If measurement is performed for 1 hour with a gate time of 10 ms then a total of 360 000 data items 1 hour 10 ms...

Страница 22: ...Sets the graph display scale automatically or resets the zoom display View Option Option Specifies the vertical axis scale View Cursor Graph Cursor Changes to cursor mode and shows a cursor line Moni...

Страница 23: ...of making a backup under the condition that you respect and obey all terms and conditions of this product license agreement 5 Prohibited actions You must not perform the following actions However if y...

Страница 24: ...The product was used for a purpose other than the original purpose The product was serviced or modified electrically or mechanically by someone other than our service personnel or authorized personnel...

Страница 25: ...nse range 320 to 900 nm Peak sensitivity wavelength 440 nm Photon detection efficiency 1 400nm 0 5p e 20 Dark count 0 5p e 500 900 kcps Temperature stability of analog output voltage 25 10 C 2 5 5 Ana...

Страница 26: ...to 900 nm Peak sensitivity wavelength 440 nm Photon detection efficiency 1 400nm 0 5p e 35 Dark count 0 5p e 600 1200 kcps Temperature stability of analog output voltage 25 10 C 2 5 5 Analog output vo...

Страница 27: ...to 900 nm Peak sensitivity wavelength 440 nm Photon detection efficiency 1 400nm 0 5p e 45 Dark count 0 5p e 900 1500 kcps Temperature stability of analog output voltage 25 10 C 2 5 5 Analog output vo...

Страница 28: ...27 B Dimensional outline Unit mm C10507 11 025U C10507 11 050U C10507 11 100U...

Страница 29: ...28 C10507 11 025C C10507 11 050C C10507 11 100C...

Страница 30: ...29 C10751 series...

Страница 31: ...tates and or other countries Other company and product names mentioned herein may be the trademarks or registered trademarks of their respective owners HAMAMATSU PHOTONICS K K Solid state division 112...

Отзывы: