29
7.7.4.2 ADVANCED SETTINGS Mode – MULTI-COOLED RF2 SETTINGS State
The following screen appears when MULTI-COOLED RF2 SETTINGS State is selected from the ADVANCED SETTINGS Mode:
• This screen allows adjustment of the IMPEDANCE CUTOFF MULTIPLIER in AUTO TEMP mode.
• The END OF RAMP IMPEDANCE is multiplied by the IMPEDANCE CUTOFF MULTIPLIER to determine the impedance value that will generate Warning (W221) and Discontinue
a Probe (RF OFF) for RISING IMPEDANCE. Refer to the EXAMPLE shown in Figure 7-25b for details.
• If the IMPEDANCE CUTOFF MULTIPLIER is “DISABLED”, then Warning (W221) is generated at 1000Ω.
Figure 7-25b ADVANCED SETTINGS Mode — MULTI-COOLED RF2 SETTINGS Display
Settings
Affected Mode
Parameter
DESCRIPTION
Range
Default Value
Increment
MULTI-COOLED RF
IMPEDANCE
CUTOFF MULTIPLIER
The IMPEDANCE CUTOFF MULTIPLIER is multiplied
by the END OF RAMP IMPEDANCE to determine the
impedance value that will generate Warning (W221)
and Discontinue a Probe (RF OFF) for RISING
IMPEDANCE.
DISABLE,
2.0 – 4.0
2.0
0.1