STATE
PARAMETER
Measurement Window
Available Modes
Available Settings
Status
Description
Status
Description
Status
Description
VOLTAGE
STIMULATION
CURRENT
STIMULATION
ON
Impedance
measurements
are displayed.
None
Other modes are
not selectable
while in the ON
state, as indicated
by the greyed-out
mode key labels.
SELECT
PROBE**
Range
A, B, C or D
Increment
N/A
STIM RATE**
Range
1-Shot –200 Hz
Available Values
†
1-Shot, 2, 5, 10, 20, 50, 75, 100, 150, 180 and 200 Hz
PULSE DUR
Range
0.1 – 1.0 ms
Available Values
†
0.1, 0.2, 0.5 and 1.0 ms
VOLTAGE/
CURRENT
Range
0.00–10.0 V/mA
Increment
0.01 V/mA
AUTO TEMP
LESION
All
measurements
are displayed.
SET TEMP
Range
38° – 95°C
Increment
1°C
TIME
Range
15 – 600 s
Increment
1 s
MANUAL
POWER
LESION
TEMP LIMIT
Range
38° – 95°C
Increment
1°C
TIME
Range
15 – 600 s
Increment
1 s
POWER LIMIT
Range
N/A
Increment
N/A
AUTO PULSED
LESION
SET TEMP
Range
38° – 95°C
Increment
1°C
TIME
Range
10 – 900 s
Increment
1 s
PULSE DUR
Range
10 – 100 ms
Increment
10 ms
PULSE RATE
Range
1 – 10 Hz
Increment
1 Hz
MANUAL
PULSED
LESION
VOLTAGE
Range
0 – 55 V
Increment
5 V (between 0 and 30 V); 1 V (greater than 30 V)
WARNING
TEMP
Range
38° – 95°C
Increment
1°C
TIME
Range
10 – 900 s
Increment
1 s
PULSE DUR
Range
20 – 100 s
Increment
10 ms
PULSE RATE
Range
1 – 10 Hz
Increment
1 Hz
** STIM RATE may not be changed until Voltage/Current set to 0 V/mA or until Output OFF; SELECT PROBE may not be changed until Output OFF.
†
Available values may be limited to two frequencies if option selected in STIMULATION SETTINGS state, accessible from the ADVANCED SETTINGS mode.