STATE
PARAMETER
Measurement Window
Available Modes
Available Settings
Status
Description
Status
Description
Status
Description
COOLED RF
TREATMENT
ON
All measurements are displayed.
None
Other modes
are not
selectable
while in the
ON state, as
indicated by
the greyed-
out mode
key labels.
SET
TEMP
Range
30° – 90°C
Increment
1°C
TIME
Range
15 s – 30 min
Increment
15 s (between 15 s and 5 min)
30 s (between 5 min and 30
min)
TRANSDISCAL
TREATMENT
SET
TEMP
Range
30° – 90°C
Increment
1°C
TIME
Range
†
1 – 30 min
Increment
30 s
RFA MODE -
TREATMENT
All measurements are displayed. If the Thermocouple Probe
is disabled via RFA SETTINGS state, accessible from the
ADVANCED SETTINGS mode, the measurement is greyed out
and the value replaced with dashed lines.
SET
TEMP
Range
38° – 95°C
Increment
1°C
TIME
Range
1 – 30 min
Increment
30 s
IDL MODE
All measurements are displayed.
During ON state, SET TEMP should not be
decreased below INITIAL TEMP in ADVANCED
SETTINGS.
During ON state, SET TIME should not be decreased
below INITIAL TIME in ADVANCED SETTINGS.
SET
TEMP
Range
40° – 95°C
Increment
1°C
TIME
Range
30 s – 20 min
Increment
30 s
†
Available values may be limited to two frequencies if option selected in STIMULATION SETTINGS state, accessible from the ADVANCED SETTINGS mode.
7.5 DONE
States
Displays and Indicators
PARAMETER
STATUS
DESCRIPTION
Available Modes
HOLD DISPLAY
Other modes are not selectable while in the DONE or ON state, as indicated by the greyed-out mode key labels.
In the DONE state, SAVE SETTINGS is replaced with HOLD DISPLAY, which has the function of holding the display. If selected, the HOLD
DISPLAY button will read RELEASE DISPLAY and the button will flash. If the flashing RELEASE DISPLAY button is selected, the system
shall return to the READY state. If the HOLD DISPLAY button is not selected, the system defaults to the ready state in 5 seconds. Settings
are not adjustable while in the DONE state, the HOLD DISPLAY state, or the RELEASE DISPLAY state.
7.6 SAVE SETTINGS State
Figure 7-21 Save Settings Display