Displays and Indicators
PARAMETER
STATUS
DESCRIPTION
RF/Stimulus Output
Stimulus Output ON
Stimulus pulses are delivered during the ON state.
RF Output ON
RF power is delivered during the ON state. SET TEMP/TEMP LIMIT/WARNING LIMIT and treatment TIME remain modifiable
during RF power delivery.
COOLED RF AND TRANSDISCAL — RF power is delivered during the TREATMENT ON state. If the procedure is stopped while
in TREATMENT ON state, the Generator enters POST-TREATMENT COOLING state if enabled in ADVANCED SETTINGS mode.
FAULT Indicator
OFF
The FAULT indicator is used only in the POST and SYSTEM FAULT states.
RF/Stimulus Output
ON/OFF Indicator
ON
The indicator is illuminated during Stimulus pulse/RF power delivery.
RF Audio Output
ON
RF audio output is generated for RF power delivery.
A tone is generated for two seconds at the conclusion of the procedure.
None
RF audio output is not generated during Stimulus pulse delivery.
A tone is generated for two seconds at the conclusion of the procedure.
Impedance Related
Audio Output
OFF
Impedance related audio output is not generated during the ON state.
Graph Window
ON
Data is displayed. The time axis (
) is indefinitely scalable during the ON state, accommodating changes to treatment
TIME. The dashed line denoting SET TEMP/TEMP LIMIT/WARNING TEMP will move according to changes made to the SET
TEMP/TEMP LIMIT/WARNING TEMP value.
COOLED RF – graphed temperatures should be the maximum Cooled RF probe temperature. Pump Status is displayed at
the top of the Graph Window.
TRANSDISCAL – graphed temperatures should be the maximum TRANSDISCAL temperature and the maximum peripheral
disc temperature, unless modified in TRANSDISCAL SETTINGS state, accessible from the ADVANCED SETTINGS mode. Pump
Status is displayed at the top of the Graph Window.
STIMULATION ON
display
The waveform “flashes” to indicate the system is generating stimulus pulses.
Status Window
None
N/A