Instruction Manual REG 21
22
8
Technical data
Measurement data
measurement ranges
0 ... 50 Pa to
0 ... 100 kPa;
see rating plate
measurement principle
inductive
change in volume
0.1…0.3 mL
overload capacity
200x (for measurement ranges < 2.5 kPa)
600 kPa (for measurement ranges > 2.5 kPa)
linearity
± 1%
± 0.5 % (optional for measurement ranges > 250 Pa)
temperature-dependent drift in
measured values
0.04% / K (+10° to +50°C)
temperature-dependent drift in the
zero point
0.04% / K (+10° to +50°C)
zero-point drift over time
0.5 %/year
Ambient conditions
medium
air, all non-aggressive gases
operating temperature
+10° to +50°C
storage temperature
-10° to +70°C
relative humidity
0 ... 80 %
EMC standards
EN 55011; EN 61000-4-3, EN 61000-4-6
conformity
declaration of conformity available upon
request
Electrical data
power consumption
3 VA
supply voltage options
standard
230 V-AC, +6 % / -15 % (50...60 Hz)
(optional)
24 V-AC, 115 V-AC, +6 % / -
15 % (50…60 Hz)
(optional)
24 V-DC, +20 % / -15 %
analogue output
standard
0..10 V (RL
≥
2 k
Ω
)
(optional)
-5 ... 5 V; 0 ... 20 mA; 4 ... 20mA
time constant
1 sec. (standard), 2.5 and 10 s may be set by user
switching outputs
standard
2 floating changeover relays; up to 230 V-Ac (50/60
Hz), 6 A w/ ohmic load
(optional)
2 bipolar NPN transistors; U
CE
< 50 V; I
C
< 200 mA,
floating
time delay to analog output
none / 20 ms (standard)
electrical connections
12-pin terminal strip
connected cross-sections
0.5 ... 2 mm
2