Instruction Manual REG 21
17
6.1
Setting parameters
If parameter mode is not released within 30 s of calling up a parameter
setting function, the instrument returns to monitoring and display mode.
Press either of the set keys (
or
) to exit parameter mode while
setting parameters (prior to concluding the session by depressing the enter
key
).
The instrument will return to monitoring and display mode if no parameters
are entered within 30 s of entering parameter mode.
The original parameters will not be affected in either of these cases.
While in parameter mode, the instrument will continue to monitor pressure and
control the switching outputs and analogue outputs.
WARNING!
Hysteresis is critical for all switching values (see fig. 5) in order to
prevent the relay from “fluttering” or the transistors from periodically switching
on/off.
6.1.1 Overview of parameter setting functions and parameters
Parameter
Parameter ID
Options
Default
values
basic settings:
depress both keys;
hold for <2 s*
time constant
n 1, n 2, n 5, n 10
F 1, F 2, F 5, F 10
1, 2, 5, 10: time constants in s
n: time constants apply to display,
analogue output and switching
outputs
F: time constants apply to display
and analogue output; switching
outputs will be triggered 20 ms
later
F 1
brightness
unitles
s value of 1…15
5
zero-point
calibration
AU: automatic zero-point
calibration
nor: manual zero-point calibration
nor
switching output 1
switch-on
value
entire measurement range
35 %
hold for >2 s**
parameter
function