4: Sample Experiments
Where:
S
λ
= the sample intensity at wavelength
λ
D
λ
= the dark intensity at wavelength
λ
R
λ
= the reference intensity at wavelength
λ
The following figure shows a typical transmission setup. The light source sends light via the input leg of a
transmission probe into a container. The light interacts with the sample. The output leg of the
transmission probe carries the information to the spectrometer, which transmits the information to the PC.
Typical Transmission Setup
Common transmission applications include measuring light through solutions, optical filters, optical
coatings, and other optical elements (such as lenses and fibers).
►
Procedure
Perform the following steps to take a transmission measurement using OOIBase32:
1. Place OOIBase32 in Scope mode by clicking the
Scope Mode
icon on the toolbar or by selecting
Spectrum | Scope Mode
from the menu bar.
2. Ensure that the entire signal is on scale. The intensity of the reference signal should peak at about
14,000 counts. If necessary, adjust the integration time until the intensity is approximately 14,000
counts.
3. Place a sample of the solvent into a cuvette and take a reference spectrum. You must take a
reference spectrum before measuring transmission.
Note
Do not put the sample itself in the path when taking a reference spectrum, only the
solvent.
220-00000-000-02-0605
21
Содержание Ocean Optics QE65000
Страница 6: ...Table of Contents iv 220 00000 000 02 0605 ...
Страница 14: ...1 Introduction 6 220 00000 000 02 0605 ...
Страница 20: ...2 Installing the QE65000 12 220 00000 000 02 0605 ...
Страница 24: ...3 Troubleshooting 16 220 00000 000 02 0605 ...
Страница 52: ...B Specifications 44 220 00000 000 02 0605 ...