2: Set-up and Operation
Set-up for Operation in External Power Control Mode
The Model LASER-785-IP-LAB laser is designed for operation in the Internal Power Preset mode and the
Models LASER-785-LAB-ADJ and LASER-785-LAB-ADJ-S are designed for operation through their
front panel controls. However, the laser provides you with the capability (via an external port on the back
panel) of adjusting the power remotely via computer or signal generator in External Power Control mode.
Use these set-up instructions for the External Power Control mode. See for
Operation in External Power
Control Mode
operation instructions in this mode.
►
Procedure
1.
Make sure that the laser power is not operating by pressing the
Laser On
switch and assuring that
the Laser On LED on the front panel is not lit.
2.
Set the Mode Selection switch on the rear panel to
External
(the amber Operating Mode LED
will light).
3.
Connect a BNC cable to the Control port on the rear panel.
4.
Go to
Operation in External Power Control Mode
to operate the laser in External Power Control
mode.
Operation
Read the following warning before attempting to use the laser:
WARNINGS
DO NOT LOOK DIRECTLY INTO THE LASER. LASER RADIATION IS
HAZARDOUS TO THE EYES.
AVOID DIRECT EXPOSURE TO THE BEAM.
Operation in Internal Power Preset Mode
Use this procedure to operate the Model LASER-785-IP-LAB laser in the Internal Power Preset mode.
►
Procedure
1.
Turn the master Power key (on the rear panel) 90 degrees clockwise to the horizontal “ON”
position. The green
System Power
LED (on the front panel) lights indicating that system power
is on.
8
LASER-785-LAB-02-0409
Содержание Ocean Optics LASER-785-IP-LAB