2: Set-up and Operation
Set-up for Operation in Internal Power Preset Mode
Use these set-up instructions for the Internal Power Preset mode (Model LASER-785-IP-LAB).
►
Procedure
1.
Make sure that the master Power key is in the vertical (“Off’) position.
2.
Connect the 120V AC power plug to the power connector on the rear of the unit.
3.
Plug the other end of the power cord into a standard 110/120V electrical outlet.
4.
Connect a FC/PC Multi-mode fiber patch cord to the FC/PC bulkhead connector on the front
panel of the unit.
5.
Note that this laser is configured with a remote interlock on the rear panel. The laser has an RJ-11
(4-pin telephone jack) that is prewired to provide a closed loop connection for the interlock.
Removal of the RJ-11 connector or a break in the closed loop circuit provided will result in a
laser interrupt and the system will be reset to standby mode. You can reconfigure the RJ-11
connector to provide a closed loop that interactively couples with their system or facility by
replacing the short length of wire with longer closed loop connections, if desired.
6.
Ensure that the
Emergency Shut-off
switch (EMO) is in the “ready” position by turning the red
button clockwise until it pops out slightly.
7.
You are now ready to turn on the laser module. Go to
Operation in Internal Power Preset Mode
.
6
LASER-785-LAB-02-0409
Содержание Ocean Optics LASER-785-IP-LAB