2: Set-up and Operation
Set-up for Operation to Control Power Through the Front
Panel
Use these set-up instructions when you want to control the laser power through the unit’s front panel
(Models LASER-785-LAB-ADJ and LASER-785-LAB-ADJ-S).
►
Procedure
1.
Make sure that the master Power key is in the vertical (“Off’) position.
2.
Connect the 120V AC power plug to the power connector on the rear of the unit.
3.
Plug the other end of the power cord into a standard 110/120V electrical outlet.
4.
Connect an SMA 905 connectorized multi-mode fiber patch cord to the SMA bulkhead connector
on the front panel of the unit.
5.
Note that this laser is configured with a remote interlock on the rear panel. The laser has an RJ-11
(4-pin telephone jack) that is prewired to provide a closed loop connection for the interlock.
Removal of the RJ-11 connector or a break in the closed loop circuit provided will result in a
laser interrupt and the system will be reset to Standby mode. You can reconfigure the RJ-11
connector to provide a closed loop that interactively couples with their system or facility by
replacing the short length of wire with longer closed loop connections, if desired.
6.
Ensure that the
Emergency Shut-off
switch (EMO) is in the “ready” position by turning the red
button clockwise until it pops out slightly.
7.
You are now ready to turn on the laser module. Go to
Operation Controlling Power Through the
Front Panel
.
LASER-785-LAB-02-0409
7
Содержание Ocean Optics LASER-785-IP-LAB
Страница 6: ...Table of Contents ii LASER 785 LAB 02 0409...
Страница 8: ...About This Manual iv LASER 785 LAB 02 0409...
Страница 12: ...1 Introduction 4 LASER 785 LAB 02 0409...
Страница 20: ...2 Set up and Operation 12 LASER 785 LAB 02 0409...
Страница 22: ...Index 14 LASER 785 LAB 02 0409...