2: Set-up and Operation
2.
Push the
Laser On
switch (on the front panel). The red LED on the front panel lights to indicate
that the laser is in operation. The laser turns on approximately 2 seconds after the switch is
pressed.
Note
Note that the switch returns to the center position after you press it. At this point, it can
function as a manual reset.
3.
In the event of a power interruption, power failure or interlock break, the laser is automatically
disabled. To re-enable it, toggle the
Laser On
switch to back to the ON position.
Operation Controlling Power Through the Front Panel
Use this procedure to control the laser power through the unit’s front panel (Models LASER-785-LAB-
ADJ and LASER-785-LAB-ADJ-S).
►
Procedure
1.
Turn the master Power key (on the rear panel) 90 degrees clockwise to the horizontal “ON”
position. The green System Power LED (on the front panel) lights indicating that system power is
on.
2.
Turn the laser adjustment dial on the front panel to the desired drive current setting. The setting is
displayed on the front panel LED.
3.
Push the
Laser On
switch (on the front panel). The red LED on the front panel lights to indicate
that the laser is in operation. The laser turns on approximately 2 seconds after the switch is
pressed.
LASER-785-LAB-02-0409
9