background image

Device Setup 

January 2023 

• 

Rev. 0 

• Standard Devices User Manual

 28

their measurement in the center of the display. If this 

option is set to 

Show Actions Menu

, an option to 

change that parameter’s engineering units is shown, 

as well as an option to highlight the parameter.

• 

Show Valve Drive

 shows or hides the valve’s 

drive percentage on controllers. See 

page 20

.

• 

Top Left Key Value

 configures which type of 

pressure (barometric, gauge, absolute) is displayed 

when an optional barometer is installed in the device.

Enable Min/Max Screen 

MENU 

→ 

SETUP 

 Display 

 Data 

Screens 

 Min/Max Screen

• 

Show 

enables the min/max screen.

• 

Hide

 disables the min/max screen.

Totalizer 1 and Totalizer 2 Screen Options

MENU 

→ 

SETUP 

 Display 

 Data Screens 

 Totalizer 1 or Totalizer 2

The totalizer screen options behave in the same manner 

as outlined in the 

Totalizer Display 

section (

page 26

). 

Performing changes in this menu performs the same change in 

the totalizer display menu. The opposite is true as well where 

changes in the totalizer display menu is reflected in this menu.

•  Flow and Batch (Default):

 View as described in 

the Totalizer Display (

page 26

), with options 

for batching, flow averaging, and  setpoint.

•  Detailed Flow: 

Similar to Flow and Batch, but 

without batching. Provides information focuses 

more on flow rates, total flow, and average flow.

•  Multi-variable: 

Screen that resembles 

the live screen (

page 11

).

Screen Lighting

MENU 

→ 

SETUP 

 Display 

 Display Lighting

The options and wording in the 

screen lighting menu

 varies 

between monochrome and color displays.

• 

On monochrome displays, press 

LESS CONTRAST

 or 

MORE CONTRAST

 to adjust the contrast levels. 

POWER 

UP Lit

 or 

Dark

 toggles whether the backlight of 

the unit will be on when the device powers on.

• 

On color displays, press 

LESS BRIGHT 

or 

BRIGHTER 

to adjust the brightness level.

Display Rotation

MENU 

→ 

SETUP 

 Display 

 Display Rotation

The device has the option of inverting (flipping) the screen 

upside-down, as configured in this menu. Use this if the device 

is installed upside down.

Advanced Setup

MENU 

 SETUP 

 Advanced

The 

advanced setup menu

 contains useful settings and 

information for troubleshooting with customer support.

Factory Restore

MENU 

→ 

SETUP 

 Advanced 

 Factory Restore

If something is not acting as expected, please contact an 

applications engineer prior to doing a 

Factory Restore 

to 

help confirm a restore is necessary. This immediately prompts 

a confirmation screen. Upon confirmation, all settings and 

registers are returned to their default settings.

 !

 Caution:

 Performing a factory restore removes 

any and all 3rd party calibrations.

Register Status

MENU 

→ 

SETUP 

 Advanced 

 Register Status

The 

Register Status

 screen displays live values for the 

internal device registers. Many of these values can help an 

applications engineer diagnose operational issues during 

technical support. Some register values clearly distinguish 

between hardware and operational problems, which speeds 

up the troubleshooting process. 

Edit Register and Device Properties

MENU 

→ 

SETUP 

 Advanced 

 Edit Register

MENU 

→ 

SETUP 

 Advanced 

 Device Properties

Editing registers and device properties allows technical 

support to fine tune device functions. These functions are 

best left alone, and used only when working with Alicat to 

diagnose or correct an issue. If your device is not functioning 

as expected, please contact Alicat support for assistance 

(

page 2

).

 

!

 

Caution:

 Editing these settings may cause the 

device to become inoperable. Do not modify them 

without working with an applications engineer.

Options for pressing buttons 

on the 

Live Screen

.

Monochrome contrast menu

.

Содержание ALICAT SCIENTIFIC M Series

Страница 1: ...OPERATING MANUAL FOR STANDARD FLOW AND PRESSURE DEVICES Models M MC P PC L LC...

Страница 2: ...d the Limited Lifetime Warranty Fill out the Service Request Form at alicat com service or contact us directly when it is time to send in your device for recalibration For devices ordered with CSA ATE...

Страница 3: ...rol Loop 20 Controlled Variable 20 PD PDF or PD I Control Algorithms 21 Flow Limit while Controlling Pressure Flow Controllers 21 Control Deadband for Pressure Control 22 Device Setup 23 Gas Selection...

Страница 4: ...7 Maintenance 38 Cleaning 38 Repair 38 Recalibration 38 Appendix A Pinouts 39 8 Pin Mini DIN Default 39 Locking Industrial Connector Pinout 39 9 Pin D Sub Connector Common Pinouts 40 15 Pin D Sub Conn...

Страница 5: ...ight page 11 Change your STP to match any standard temperature and pressure reference for your mass flow page 25 Log data to your PC with a serial data connection to control the device and capture dat...

Страница 6: ...SETUP Sensor Totalizer 1 or Totalizer 2 See page 25 Connectors and Buttons The drawing to the right represents a typical configuration of a standard controller Your device s appearance and connection...

Страница 7: ...and process connection if present For pressure controllers refer to Appendix C page 45 for diagrams on plumbing specific controllers Face seal fittings do not need Teflon tape applied to the threads...

Страница 8: ...5 12 Vdc Output as the standard 0 5 Vdc analog output signal Pin 2 is normally a constant 5 12 Vdc Controller 8 Pin Mini DIN Pinout Meter and Gauge 8 Pin Mini DIN Pinout Pin Function Pin Function 1 No...

Страница 9: ...r gauge remotely by momentarily grounding pin 4 When the switch is closed the device tares Operation resumes when the switch is released You can also tare with the front controls page 12 or serial com...

Страница 10: ...the upstream bleed port screw 1 to 2 turns or until liquid begins to leak from the threads Do not remove the screw as it has pressure behind it is very small easy to lose and delicate to rethread 2 G...

Страница 11: ...se button 5 Pressure and liquid devices do not measure mass flow Highlight temperature mass flow and liquid devices use button 2 Pressure devices do not measure temperature Changing the setpoint contr...

Страница 12: ...W BEFORE PRESSING TARE may display on some devices If so press TARE to confirm taring the flow Taring Pressure TARE PRESS or MENU TARES TARE PRESS Taring pressure requires the device to be open to atm...

Страница 13: ...anti corrosive line of mass flow devices As soon as you press SET from the gas list your device reconfigures its flow rate calculations to the newly selected gas properties There is no need to restart...

Страница 14: ...17 for setpoint instructions NEXT moves to the totalizer screen if enabled or MENU opens the device menu Totalizer Screens Flow Devices MENU SETUP Sensor Totalizer Totalizer 1 or Totalizer 2 Mass flow...

Страница 15: ...tions RED Parameters are displayed in red when its value exceeds 128 of the device s specifications YELLOW Menu items that are ready to be selected appear in yellow The color replaces the symbol in se...

Страница 16: ...Initials of the person who calibrated the device SW Firmware version Device Full Scale Ranges MENU ABOUT Full Scale Ranges This displays the maximum calibrated range of available flow and pressure re...

Страница 17: ...e taring Delay Before Tare manages how much time the controller waits before taring Make sure the delay provides enough time for the process to stop flow and settle Caution Auto tare typically is not...

Страница 18: ...tpoint Limits The setpoint limits menu configures upper and lower limits for selecting a setpoint By default the limits are the controller s measuring range but more strict limits may be beneficial in...

Страница 19: ...t volume of gas flows through the controller the valve closes and flow stops You can repeat batches with a single button press To utilize batching at least one totalizer must be enabled The BATCH butt...

Страница 20: ...n the main display and or as part of transmitted serial data There are four valve display options None No valve information is displayed Main Screen Only on the main display Digital Data Only in the s...

Страница 21: ...ill predict needed future corrections based on the current rate of change in the system This often results in slowing the system down to minimize overshoot and oscillations Troubleshooting Valve Perfo...

Страница 22: ...OL Control Deadband Deadband The controller must first reach the setpoint for the deadband to engage If the process variable drifts outside a deadband limit active control resumes until the setpoint i...

Страница 23: ...erties and exits to the setup menu COMPOSER Gas Mixes MENU SETUP Active Gas COMPOSER Mixes To remain accurate your mass flow device needs to reference the viscosity of the gas you are flowing through...

Страница 24: ...a second page use the PAGE button to see the remaining list Viewing Deleting and Creating Similar Mixes MENU SETUP Active Gas COMPOSER Mixes Select mix INFO The current configuration of any existing C...

Страница 25: ...troller with a zero band value of 0 25 displays as 0 SLPM for all readings below 0 05 SLPM This function also applies to gauge pressure readings when using the optional barometer Totalizer Options Flo...

Страница 26: ...lizer Digital Data MENU SETUP Sensor Totalizer Totalizer 1 or Totalizer 2 Digital Data This setting enables the device to report the totalizer value whenever it is polled by the serial terminal page 2...

Страница 27: ...d can be set in seconds up to 99999 9 seconds 1 day 3 hours 46 minutes 39 9 seconds Baud Rate SETUP RS 232 Serial or RS 485 Serial Baud Rate Baud rate is the speed at which digital devices transfer in...

Страница 28: ...will be on when the device powers on On color displays press LESS BRIGHT or BRIGHTER to adjust the brightness level Display Rotation MENU SETUP Display Display Rotation The device has the option of in...

Страница 29: ...l Alicat devices Within its features is a terminal that allows for direct serial communication with connected devices Flowvision 2 0 can be downloaded along with its user manual from alicat com flowvi...

Страница 30: ...nual taring can be accomplished through two separate commands for flow and pressure Taring flow sets the zero flow reading and must be done when no flow is passing through the device Tare flow unit_id...

Страница 31: ...POSER gas mixes have a mix number between 236 and 255 Choose a user mix unit_idGGas Number Example AG255 reconfigures for user mix 255 Defining a new COMPOSER gas mix is faster using serial commands t...

Страница 32: ...red_unit_id Set streaming interval unit_idNCS number_of_ms Query live data info unit_id D Manufacturer info unit_id M Firmware version unit_idve Lock the front display unit_idL Unlock the display unit...

Страница 33: ...RTU Serial Protocol Alicat uses the Modbus standard of offsetting registers by 1 from addresses meaning register 1 is equivalent to address 0 However some systems expect data to be mapped as 0 indexed...

Страница 34: ...send a command to register 1000 without sending an argument to 1001 the device will interpret the command with a default argument of 0 At any time after sending a special command to registers 1000 10...

Страница 35: ...Controllers 0 65535 10 Change I in PID Loop 0 Mass Flow Controllers 1 Mass Flow and Liquid Controllers 2 PSID Controllers 3 Abs Press Controllers and Mass Flow Controllers 4 Gauge Pressure Controller...

Страница 36: ...ering Units From this menu you can adjust any variable s units For more information see page 24 Issue My controller won t reach its setpoint Action This is often caused by not enough supply pressure i...

Страница 37: ...o check before changing gases or liquids is the chemical and material compatibility of the gas being used with the wetted materials inside the controller We also recommend contac ting suppor t page 2...

Страница 38: ...device can be cleaned with a soft dry cloth Warning If you suspect that debris or other foreign material has entered your device do not take apart the flow body to clean it Please contact support for...

Страница 39: ...Serial RS 232 RX RS 485 Input Signal receive 3 Serial RS 232 RX RS 485 Input Signal receive 4 Analog Setpoint Input Controllers Remote Tare Meters and Gauges 4 Analog Setpoint Input Controllers Remot...

Страница 40: ...log Out RX or A Analog Out NC Analog Out 3 Ground Analog In NC NC NC Ground NC 4 NC Ground NC NC NC Analog Out NC 5 Power In Ground Power In Ground Ground Power In Ground 6 NC Ground NC Analog Out NC...

Страница 41: ...t Input Controllers Ground to Tare Meters and Gauges Analog Setpoint Input Controllers Ground to Tare Meters and Gauges 5 Serial RS 232TX RS 485 output signal send Serial RS 232TX RS 485 output signal...

Страница 42: ...r L s SL s NL s Liter per second LPM SLPM NLPM Liter per minute L h SL h NL h Liter per hour US GPM US gallon per minute US GPH US gallon per hour CCS SCCS NCCS Cubic centimeter per second CCM SCCM NC...

Страница 43: ...2 150 Bio30M 30 CH4 70 CO2 151 Bio35M 35 CH4 65 CO2 152 Bio40M 40 CH4 60 CO2 153 Bio45M 45 CH4 55 CO2 154 Bio50M 50 CH4 50 CO2 155 Bio55M 55 CH4 45 CO2 156 Bio60M 60 CH4 40 CO2 157 Bio65M 65 CH4 35 CO...

Страница 44: ...Deuterium D2 Ethane C2H6 Ethylene Ethene C2H4 Helium He Hydrogen H2 Krypton Kr Methane CH4 Neon Ne Nitrogen N2 Nitrous Oxide N2O Oxygen O2 Propane C3H8 Sulfur Hexafluoride SF6 1 Xenon Xe Breathing Gas...

Страница 45: ...lenoid is energized pressure is delivered to the shutoff valve causing it to open When the solenoid is returned to a relaxed state the gas vents to atmosphere allowing the shut off valve to close Pres...

Страница 46: ...re Differential Pressure Control Differential pressure controllers have two ports for connecting to points in the system where the differential pressure is measured The upstream port is for the higher...

Страница 47: ...menus available on the screen Integration of the vacuum sensor control valve and PID algorithm in a single device eliminates the need for an external vacuum sensor and throttle valve in your system T...

Отзывы: