background image

Process Control 

January 2023 

• 

Rev. 0 

• Standard Devices User Manual

 18

Idle Connection Response

MENU 

 CONTROL 

 Setpoint Setup

 → 

On Comm Timeout

If the communication connection is timed out, the controller 

can either set a zero setpoint, or maintain the last setpoint 

given. The timeout time is infinite by default and can be 

manually defined (

page 27

).

Setpoint on Power-Up

By default, the controller remembers its last setpoint across 

power cycles, but a specific setpoint can be configured to be 

used on every power-up. It is possible to also configure the 

power-up setpoint to follow the ramp rate. 

Power-Up Setpoint Value

MENU 

 CONTROL 

 Setpoint Setup 

 

Power Up Setpoint 

 Value

By selecting 

Fixed Setpoint

 and entering a value, the 

device moves to the same setpoint every time after power-up, 

ignoring whatever the setpoint was before being powered off.
If the setpoint is digitally provided more than every few 

minutes, use a fixed setpoint on power-up to avoid wearing 

out non-volatile memory in the device.

Power-Up Setpoint with Ramping

MENU 

 CONTROL 

 Setpoint Setup 

 

Power Up Setpoint 

 Ramp

This designates if the device follows the ramp rate or not after 

powering on. The device can either honor the ramp rate and 

ramp up (

Honor from 0

) or jump immediately to the power-up 

setpoint (

Jump from 0

) . 

For more information on ramping, see 

Setpoint Ramping 

(

page 18

). This functions the same as the power-up option 

in 

Ramping Options

. Changing one overwrites the other.

Setpoint Limits

MENU 

 CONTROL 

 Setpoint Setup

 → 

Setpoint Limits

The 

setpoint limits menu

 configures upper and lower 

limits for selecting a setpoint. By default, the limits are the 

controller's measuring range, but more strict limits may be 

beneficial in certain applications.

Over a serial connection, the controller rejects requests of 

a setpoint outside the limit and an error is returned. When 

using an analog setpoint signal, setpoints that are outside of 

the setpoint limits are treated as if they were at the nearest 

limit. For example, if you request a setpoint via analog that is 

below the lower limit, the controller sets the setpoint at the 

lower limit.

 

!

 

Caution:

 Flow controllers that have non-zero 

lower setpoint limits cannot be set to stop flow 

until the lower limit has been cleared.

 ✓

 Note:

 When changing from one control loop variable 

to another, the controller remembers setpoint limits 

as percentages of full scale. For example, a 10-SLPM 

limit on a 20-SLPM controller (50% of full scale) will 

become a limit of 80 PSIA (50% of 160 PSIA) if the 

control loop is changed to absolute pressure.

Setpoint Ramping

MENU 

 CONTROL 

 Setpoint Ramp

Setpoint ramping regulates how quickly the controller reaches 

a new setpoint. It is often used to prevent bursts of pressure 

or flow from damaging delicate instruments at the start of a 

process.
To activate setpoint ramping, set a maximum ramp rate and 

configure when to enable the ramping function.

Ramp Rate

• 

Ramp

 changes the maximum rate of change. 

• 

Units

 changes the engineering units used.

• 

Set By Delta / Time

 allows for more 

detailed control of the ramp rate including 

changing the value of the time period.

Setting a maximum ramp rate.

Configuring delta / time ramping.

S

etpoint ramping menu

.

Power-up setpoint menu

.

Setting a minimum setpoint.

Setpoint limits menu

.

Communication timeout menu

.

Содержание ALICAT SCIENTIFIC M Series

Страница 1: ...OPERATING MANUAL FOR STANDARD FLOW AND PRESSURE DEVICES Models M MC P PC L LC...

Страница 2: ...d the Limited Lifetime Warranty Fill out the Service Request Form at alicat com service or contact us directly when it is time to send in your device for recalibration For devices ordered with CSA ATE...

Страница 3: ...rol Loop 20 Controlled Variable 20 PD PDF or PD I Control Algorithms 21 Flow Limit while Controlling Pressure Flow Controllers 21 Control Deadband for Pressure Control 22 Device Setup 23 Gas Selection...

Страница 4: ...7 Maintenance 38 Cleaning 38 Repair 38 Recalibration 38 Appendix A Pinouts 39 8 Pin Mini DIN Default 39 Locking Industrial Connector Pinout 39 9 Pin D Sub Connector Common Pinouts 40 15 Pin D Sub Conn...

Страница 5: ...ight page 11 Change your STP to match any standard temperature and pressure reference for your mass flow page 25 Log data to your PC with a serial data connection to control the device and capture dat...

Страница 6: ...SETUP Sensor Totalizer 1 or Totalizer 2 See page 25 Connectors and Buttons The drawing to the right represents a typical configuration of a standard controller Your device s appearance and connection...

Страница 7: ...and process connection if present For pressure controllers refer to Appendix C page 45 for diagrams on plumbing specific controllers Face seal fittings do not need Teflon tape applied to the threads...

Страница 8: ...5 12 Vdc Output as the standard 0 5 Vdc analog output signal Pin 2 is normally a constant 5 12 Vdc Controller 8 Pin Mini DIN Pinout Meter and Gauge 8 Pin Mini DIN Pinout Pin Function Pin Function 1 No...

Страница 9: ...r gauge remotely by momentarily grounding pin 4 When the switch is closed the device tares Operation resumes when the switch is released You can also tare with the front controls page 12 or serial com...

Страница 10: ...the upstream bleed port screw 1 to 2 turns or until liquid begins to leak from the threads Do not remove the screw as it has pressure behind it is very small easy to lose and delicate to rethread 2 G...

Страница 11: ...se button 5 Pressure and liquid devices do not measure mass flow Highlight temperature mass flow and liquid devices use button 2 Pressure devices do not measure temperature Changing the setpoint contr...

Страница 12: ...W BEFORE PRESSING TARE may display on some devices If so press TARE to confirm taring the flow Taring Pressure TARE PRESS or MENU TARES TARE PRESS Taring pressure requires the device to be open to atm...

Страница 13: ...anti corrosive line of mass flow devices As soon as you press SET from the gas list your device reconfigures its flow rate calculations to the newly selected gas properties There is no need to restart...

Страница 14: ...17 for setpoint instructions NEXT moves to the totalizer screen if enabled or MENU opens the device menu Totalizer Screens Flow Devices MENU SETUP Sensor Totalizer Totalizer 1 or Totalizer 2 Mass flow...

Страница 15: ...tions RED Parameters are displayed in red when its value exceeds 128 of the device s specifications YELLOW Menu items that are ready to be selected appear in yellow The color replaces the symbol in se...

Страница 16: ...Initials of the person who calibrated the device SW Firmware version Device Full Scale Ranges MENU ABOUT Full Scale Ranges This displays the maximum calibrated range of available flow and pressure re...

Страница 17: ...e taring Delay Before Tare manages how much time the controller waits before taring Make sure the delay provides enough time for the process to stop flow and settle Caution Auto tare typically is not...

Страница 18: ...tpoint Limits The setpoint limits menu configures upper and lower limits for selecting a setpoint By default the limits are the controller s measuring range but more strict limits may be beneficial in...

Страница 19: ...t volume of gas flows through the controller the valve closes and flow stops You can repeat batches with a single button press To utilize batching at least one totalizer must be enabled The BATCH butt...

Страница 20: ...n the main display and or as part of transmitted serial data There are four valve display options None No valve information is displayed Main Screen Only on the main display Digital Data Only in the s...

Страница 21: ...ill predict needed future corrections based on the current rate of change in the system This often results in slowing the system down to minimize overshoot and oscillations Troubleshooting Valve Perfo...

Страница 22: ...OL Control Deadband Deadband The controller must first reach the setpoint for the deadband to engage If the process variable drifts outside a deadband limit active control resumes until the setpoint i...

Страница 23: ...erties and exits to the setup menu COMPOSER Gas Mixes MENU SETUP Active Gas COMPOSER Mixes To remain accurate your mass flow device needs to reference the viscosity of the gas you are flowing through...

Страница 24: ...a second page use the PAGE button to see the remaining list Viewing Deleting and Creating Similar Mixes MENU SETUP Active Gas COMPOSER Mixes Select mix INFO The current configuration of any existing C...

Страница 25: ...troller with a zero band value of 0 25 displays as 0 SLPM for all readings below 0 05 SLPM This function also applies to gauge pressure readings when using the optional barometer Totalizer Options Flo...

Страница 26: ...lizer Digital Data MENU SETUP Sensor Totalizer Totalizer 1 or Totalizer 2 Digital Data This setting enables the device to report the totalizer value whenever it is polled by the serial terminal page 2...

Страница 27: ...d can be set in seconds up to 99999 9 seconds 1 day 3 hours 46 minutes 39 9 seconds Baud Rate SETUP RS 232 Serial or RS 485 Serial Baud Rate Baud rate is the speed at which digital devices transfer in...

Страница 28: ...will be on when the device powers on On color displays press LESS BRIGHT or BRIGHTER to adjust the brightness level Display Rotation MENU SETUP Display Display Rotation The device has the option of in...

Страница 29: ...l Alicat devices Within its features is a terminal that allows for direct serial communication with connected devices Flowvision 2 0 can be downloaded along with its user manual from alicat com flowvi...

Страница 30: ...nual taring can be accomplished through two separate commands for flow and pressure Taring flow sets the zero flow reading and must be done when no flow is passing through the device Tare flow unit_id...

Страница 31: ...POSER gas mixes have a mix number between 236 and 255 Choose a user mix unit_idGGas Number Example AG255 reconfigures for user mix 255 Defining a new COMPOSER gas mix is faster using serial commands t...

Страница 32: ...red_unit_id Set streaming interval unit_idNCS number_of_ms Query live data info unit_id D Manufacturer info unit_id M Firmware version unit_idve Lock the front display unit_idL Unlock the display unit...

Страница 33: ...RTU Serial Protocol Alicat uses the Modbus standard of offsetting registers by 1 from addresses meaning register 1 is equivalent to address 0 However some systems expect data to be mapped as 0 indexed...

Страница 34: ...send a command to register 1000 without sending an argument to 1001 the device will interpret the command with a default argument of 0 At any time after sending a special command to registers 1000 10...

Страница 35: ...Controllers 0 65535 10 Change I in PID Loop 0 Mass Flow Controllers 1 Mass Flow and Liquid Controllers 2 PSID Controllers 3 Abs Press Controllers and Mass Flow Controllers 4 Gauge Pressure Controller...

Страница 36: ...ering Units From this menu you can adjust any variable s units For more information see page 24 Issue My controller won t reach its setpoint Action This is often caused by not enough supply pressure i...

Страница 37: ...o check before changing gases or liquids is the chemical and material compatibility of the gas being used with the wetted materials inside the controller We also recommend contac ting suppor t page 2...

Страница 38: ...device can be cleaned with a soft dry cloth Warning If you suspect that debris or other foreign material has entered your device do not take apart the flow body to clean it Please contact support for...

Страница 39: ...Serial RS 232 RX RS 485 Input Signal receive 3 Serial RS 232 RX RS 485 Input Signal receive 4 Analog Setpoint Input Controllers Remote Tare Meters and Gauges 4 Analog Setpoint Input Controllers Remot...

Страница 40: ...log Out RX or A Analog Out NC Analog Out 3 Ground Analog In NC NC NC Ground NC 4 NC Ground NC NC NC Analog Out NC 5 Power In Ground Power In Ground Ground Power In Ground 6 NC Ground NC Analog Out NC...

Страница 41: ...t Input Controllers Ground to Tare Meters and Gauges Analog Setpoint Input Controllers Ground to Tare Meters and Gauges 5 Serial RS 232TX RS 485 output signal send Serial RS 232TX RS 485 output signal...

Страница 42: ...r L s SL s NL s Liter per second LPM SLPM NLPM Liter per minute L h SL h NL h Liter per hour US GPM US gallon per minute US GPH US gallon per hour CCS SCCS NCCS Cubic centimeter per second CCM SCCM NC...

Страница 43: ...2 150 Bio30M 30 CH4 70 CO2 151 Bio35M 35 CH4 65 CO2 152 Bio40M 40 CH4 60 CO2 153 Bio45M 45 CH4 55 CO2 154 Bio50M 50 CH4 50 CO2 155 Bio55M 55 CH4 45 CO2 156 Bio60M 60 CH4 40 CO2 157 Bio65M 65 CH4 35 CO...

Страница 44: ...Deuterium D2 Ethane C2H6 Ethylene Ethene C2H4 Helium He Hydrogen H2 Krypton Kr Methane CH4 Neon Ne Nitrogen N2 Nitrous Oxide N2O Oxygen O2 Propane C3H8 Sulfur Hexafluoride SF6 1 Xenon Xe Breathing Gas...

Страница 45: ...lenoid is energized pressure is delivered to the shutoff valve causing it to open When the solenoid is returned to a relaxed state the gas vents to atmosphere allowing the shut off valve to close Pres...

Страница 46: ...re Differential Pressure Control Differential pressure controllers have two ports for connecting to points in the system where the differential pressure is measured The upstream port is for the higher...

Страница 47: ...menus available on the screen Integration of the vacuum sensor control valve and PID algorithm in a single device eliminates the need for an external vacuum sensor and throttle valve in your system T...

Отзывы: