PRESSURE CONTROLLER operating manual
19
Control Options
The control options menu allows the user to enable ramping in either
direction, establish a control deadband, or create setpoint limits.
Menu
→
Control
→
Advanced Control
→
Control Options
•
RAMP ENABLE
Defines the
conditions when setpoint ramping
will be active (
).
•
DEAD BAND
Sets the size of the
control deadband (below).
•
BACK
Returns to the Advanced
Control Menu (
•
SETPT LIMITS
Sets upper and
lower setpoint limits (
•
MAIN
Exits to the Main
Display (
).
Control Deadband
Menu
→
Control
→
Advanced Control
→
Control Options
→
Dead Band
The control deadband on the pressure
controller allows the user to select
upper and lower bounds around the
setpoint. The controller will not adjust
its valve position if the measured
pressure is within these bounds. In this
example, the controller will not adjust
its valve position if the measured pres-
sure is within 5 PSIG of the setpoint.
Control Options Menu
Control Deadband Menu