NOTE :

If solder can't wet easily on the conductive plate due to low tip temperature or because of 

the tip having small thermal capacity, use the included conductive wire.

Replacement method

Remove the 2 screws attaching the conductive plate, replace the conductive plate with the 

conductive wire, and then attach the conductive wire with the removed screws.

4. OPERATION(continued)

NOTE :

During measurement of leak voltage, even 

when the tip is not applied to the conductive 

plate, values are displayed; however, this is 

not a malfunction of the product. In addition, 

when values out of measurement range are 

detected, burnout will be displayed, and this is 

also not a malfunction.

1)  Plug the power plug of the soldering 

iron to be measured into the power 

receptacle the HAKKO FG-101B 

main body.

2)  Wait until the tip reaches the 

set temperature. If the soldering 

iron temperature is variable, set 

temperature to maximum.

3)  Press the SELECT button and 

change the mode to "

mV

".

4)  Press the 

 button.

Refer to "

 AUTO ZERO function

(page 8) in "

4-5 Explanation about 

various functions

".

5)  Clean the tip and apply a new coat of 

solder on the tip.

6)  Apply a coating of solder to the center 

of the conductive plate and heat it 

until a good soldering wet condition is 

achieved.

7)  When the display temperature is 

stable, read the values.

4-3  Measurement of leak voltage

 

CAUTION

•  Before measuring the voltage or resistance, be sure to connect the main body 

power cord to a two-pole grounded-type receptacle.

•  If values exceeding specified ones are displayed as results for voltage or 

resistance, check the tip and attaching screws of the soldering iron for looseness 

and take another measurement.

Содержание FG101B-03

Страница 1: ...emperature leak voltage and ground resistance Please read the manual carefully before operating the HAKKO FG 101B Please keep this manual readily accessible for reference Instruction Manual SOLDERING...

Страница 2: ...TIONS AND NOTES 3 4 OPERATION 4 4 1 Initial setup 4 4 2 Measurement of tip temperature 5 4 3 Measurement of leak voltage 6 4 4 Measurement of resistance difference between tip and ground wire 7 4 5 Ex...

Страница 3: ...isplay contents 1 This alarm indicates sensor burnout If this alarm occurs replace the sensor 2 When the and buttons are held pressed simultaneously for a long time 1 sec or longer the count is reset...

Страница 4: ...nctions Temperature measurement count function 000 to 999 times Power consumption 100V 3 2W 100 110V 3 6W 120V 3 2W 220 240V 3 6W Outline dimensions 211 W 53 H 126 D mm Weight 0 95 kg Usable temperatu...

Страница 5: ...rning use of the appliance in safe way and understand the hazards involved Children shall not play with the appliance Cleaning and user maintenance shall not be made by children without supervision Be...

Страница 6: ...will move toward the terminal 2 Attach the sensor with the slide pin moved toward the terminal 3 Attach the sensor having a red mark to the red colored terminal and the sensor having a blue mark to th...

Страница 7: ...When flux is stuck on the terminal wipe it off with alcohol Do not use thinner or benzin for wiping Read the values when the display temperature is stable CAUTION Do not measure the temperature by ap...

Страница 8: ...ed into the power receptacle the HAKKO FG 101B main body 2 Wait until the tip reaches the set temperature If the soldering iron temperature is variable set temperature to maximum 3 Press the SELECT bu...

Страница 9: ...the button will be displayed Pressed for a long time 1 second or longer The MAX HOLD function will be released and the product will return to normal display CAUTION Turning off the power always cance...

Страница 10: ...Each time the button is pressed the AUTO HOLD function toggles between ON and OFF The AUTO HOLD function does not assume the measurement by temperature probes such as the hot air Data send function in...

Страница 11: ...performed smoothly using the functions of connected stations Ground terminal GND When using a type of soldering iron that is grounded with an alligator clip connect the clip to the GND terminal Maint...

Страница 12: ...r cord 3 core BS plug India B2424 Power cord 3 core European plug 220V KC 230V CE B2425 Power cord 3 core BS plug 230V CE U K B2426 Power cord 3 core Australian plug B2436 Power cord 3 core Chinese pl...

Отзывы: