background image

5

CAUTION

•  Since the main body is partially 

made of resin, be careful not to 
touch the tip to the main body. In 
addition, also be careful not to 
touch the tip to the terminal or the 
slide pin.

•  Although the measurement point 

of the sensor undergoes special 
treatment, it will deteriorate 
gradually due to repeated 
measurements. In order to measure 
the temperature correctly, if the 
measurement point is worn out, 
replace the sensor with a new 
one. The reference number for 
sensor replacement is approx. 50 
measurements.

•  When flux is stuck on the terminal, 

wipe it off with alcohol. (Do not use 
thinner or benzin for wiping.)

•  Read the values when the display 

temperature is stable.

 

CAUTION

Do not measure the temperature by applying 

hot air (HAKKO FR-811, etc.) directly to the 

HAKKO FG-101B. If hot air is applied directly, 

the HAKKO FG-101B will be damaged.

4-2  Measurement of tip 

temperature

1)  Press the SELECT button to set the 

mode to "

TEMP

".

2)  Apply a new coat of solder on 

the tip and touch it on the sensor 

measurement part (See the figures at 

right.)

New solder

Measurement 
point

Tip

NOTE:

When measuring, apply a new coat of solder on the tip. This action is required to ensure 

that the conductive plate comes in firm contact with the tip.

Содержание FG101B-03

Страница 1: ...emperature leak voltage and ground resistance Please read the manual carefully before operating the HAKKO FG 101B Please keep this manual readily accessible for reference Instruction Manual SOLDERING...

Страница 2: ...TIONS AND NOTES 3 4 OPERATION 4 4 1 Initial setup 4 4 2 Measurement of tip temperature 5 4 3 Measurement of leak voltage 6 4 4 Measurement of resistance difference between tip and ground wire 7 4 5 Ex...

Страница 3: ...isplay contents 1 This alarm indicates sensor burnout If this alarm occurs replace the sensor 2 When the and buttons are held pressed simultaneously for a long time 1 sec or longer the count is reset...

Страница 4: ...nctions Temperature measurement count function 000 to 999 times Power consumption 100V 3 2W 100 110V 3 6W 120V 3 2W 220 240V 3 6W Outline dimensions 211 W 53 H 126 D mm Weight 0 95 kg Usable temperatu...

Страница 5: ...rning use of the appliance in safe way and understand the hazards involved Children shall not play with the appliance Cleaning and user maintenance shall not be made by children without supervision Be...

Страница 6: ...will move toward the terminal 2 Attach the sensor with the slide pin moved toward the terminal 3 Attach the sensor having a red mark to the red colored terminal and the sensor having a blue mark to th...

Страница 7: ...When flux is stuck on the terminal wipe it off with alcohol Do not use thinner or benzin for wiping Read the values when the display temperature is stable CAUTION Do not measure the temperature by ap...

Страница 8: ...ed into the power receptacle the HAKKO FG 101B main body 2 Wait until the tip reaches the set temperature If the soldering iron temperature is variable set temperature to maximum 3 Press the SELECT bu...

Страница 9: ...the button will be displayed Pressed for a long time 1 second or longer The MAX HOLD function will be released and the product will return to normal display CAUTION Turning off the power always cance...

Страница 10: ...Each time the button is pressed the AUTO HOLD function toggles between ON and OFF The AUTO HOLD function does not assume the measurement by temperature probes such as the hot air Data send function in...

Страница 11: ...performed smoothly using the functions of connected stations Ground terminal GND When using a type of soldering iron that is grounded with an alligator clip connect the clip to the GND terminal Maint...

Страница 12: ...r cord 3 core BS plug India B2424 Power cord 3 core European plug 220V KC 230V CE B2425 Power cord 3 core BS plug 230V CE U K B2426 Power cord 3 core Australian plug B2436 Power cord 3 core Chinese pl...