序号/
No.
0
1
2
时间/
Time
20130425
20130425
20130425
设定温度/
Setting Temp.
5
5
5
箱内温度/
Inner Temp.
5.6
5
4.9
...
...
...
...
Press the “Cal” for 10s and the display area of inner temperature shows “1P” stably. Press the “Cal ”,
and the display area of inner temperature blinks the year. Press the “Sensor ” (decrease) or
“Cal Cancel” (increase) to choose the year. It displays 10 to 99 recurrently. For example, to set 2013,
choose 13 and press the “Cal ” to save it. And then the display area of inner temperature shows "2P"
stably. Press the “Cal”, and the display area of inner temperature blinks the month . Press the
“Sensor ” (decrease) or “Cal Cancel” (increase) to choose the month. It displays 01 to 12 recurrently.
When the current month comes out, press the “Cal ” to save it. And then the display area of inner
temperature shows “3P” stably. Press the “Cal ”, and the display area of inner temperature blinks
the date . Press the“Sensor ” (decrease) or “Cal Cancel” (increase) to choose the date. It displays
01 to 31 recurrently. When the current local date comes out, press the “Cal ” to save it. After that,
the display area of inner temperature shows “4P” stably. Press the “Cal ”, and the display area of
inner temperature blinks the hour.Press the “Sensor ” (decrease) or “Cal Cancel” (increase) to
choose the hour. It displays 00 to 23 recurrently. When the current local hour comes out, press the
“ Cal ” to save it. After that, the display area of inner temperature shows “5P” stably. Press the
“ Cal ”, and the display area of inner temperature blinks the minute. Press the “Sensor ”(decrease)
or “Cal Cancel” (increase) to choose the minute. It displays 00 to 59 recurrently. When the current
local minute comes out, press the “Cal ” to save it. After that, the display area of inner temperature
USB Function
The computer has a USB function. Test data can be exported via the USB port. The computer can
automatically collect and store the test data within the recent 10 years. It collects data every other 6
minutes and stores them automatically. The latest data will automatically replace the earliest data
when the data storage is full.With a USB port plugged in, the computer will automatically identify it
and start to export data to the USB port.In the data exporting process, if you press “ Cal ” , the
screen will blink “ USb” , which indicates the data are now exporting and not finished yet. After
blinking 5s, it exits and displays the actual temperature in the case.If the displays “ ALL” stably, the
data export is finished. It exits 5s later can displays the actual temperature in the case.The USB port
can be removed at this point. The exported data are in the following formats:
Functions of USB interface
Time setting of USB interface data recording (time system current setting):
26
19
HYC-940F
Control panel
Light switch
Door lock
Door ajar alarm switch
Handle
Caster
Leveling leg