•Adjustment of Calibration of the Recorder

The  recorder  is  accurately adjusted  before  leaving the  plant. It  will  not be  affected  even  if  the 
power is cut off . If adjustment is required, please perform a calibration according  to the following 
procedures:
1. Operate the refrigerator continuously. Use the recorder to continuously record  the temperature 
curve for at least 2 hours after the refrigerator has reached a steady state (fl uctuation of displayed 
temperature is regular).
2. Put a standard thermoelectric couple into a  solution bottle with 100ml of glycerol (10%). And 
place the bottle in the box together with sensor of the recorder;
3.  Compare  the  temperature  values  of  standard  thermoelectric  couple  and  recorder  when 
the bottle  is  placed  in the  refrigerator for about  4  hours  and the  temperature  of  glycerol  is  the 
same with that in  the box. If temperature value of the recorder is  different with that  of standard 
thermoelectric couple,  press the  button on the left of the panel (1#)  or on the right side (2#)  to 
adjust the temperature value on the recorder and to make it in accordance with indicated value of 
the thermoelectric couple.
Note: The recording pen starts moving after the button is released for 5 seconds.

The recorder was calibrated before leaving the plant. Please do not adjust it in non-
special situation.

Caution

21

• USB function is (optional for HXC-158/158B/358/358B/608/608B).

The control panel has a USB port, which allows extraction of temperature data and other test data via 
a jump drive. It can automatically collect and store the test data within the recent 10 years. The latest 
data will automatically replace the earliest data when the data storage is full. Plug in a jump drive, the 
unit will automatically identify it and start to export the data into the jump drive. While data are being 
transferred, press “Minus” (“Cal”), the display panel fl ashes “USB”, which indicates the data are now 
exporting and not fi nished yet. After 5 seconds, the display will return to the temperature mode. If the 
display shows “ALL”, indicating the data export process is completed. The jump drive can be removed 
and  it  can  be  plugged  into  a  computer  to  read  the  data. The  exported  data  are  in  the  following 
formats:

No.

Time

Setting Temp.

Inner Temp.

00000

20130425 11:40

4

4.6

00001

20130425 11:46

4

4

00002

20130425 11:52

4

3.9

•Time setting of the record data (the present time setting of system):

Press and  hold the “Minus”  (“Cal”)  key for  10  seconds, “1P”  is  shown on the display. Press  “Set” 
(“Cal”) key, and the selection of year is in fl ashing mode. Press “Minus” or “Plus” (“Sensor” or “Cal 
cancel”) key to enter the current year, for example, 16, which means 2016. Press “Set” (“Cal”) to save 
and confi rm. Next “2P” is displayed, repeat the same  process to enter the month. Next repeat the 
same process to enter the date for “3P” mode. Repeat the same procedure to enter hour and minute 
under “4P” and “5P”. After all entries are completed, press and hold “Set”(“Cal” ) key for 5 seconds or 
there is no operation in 10 seconds, the new data will be saved, the control exits back to temperature 
mode.
•Note:

Among  the  setting procedures of  aforementioned 1P-5P,  parameters can  be  adjusted selectively 
through “Minus” or “Plus”( “Sensor” or “Cal cancel” ) key. For example, if the year(1P) is displayed 
in display area, and the year (1P) and month (2P) do not need to be adjusted, reset the date (3P) 
directly, you can Press “Plus” (“Cal cancel”) key to choose the date (3P), then press “Set“(“Cal”) key, 
at this time, the display area fl ashes date, then press “Minus” or “Plus”(“Sensor” or “Cal cancel” ) key 
to adjust the present date, then press “Set” ( “Cal”) key to save and confi rm. When parameter setting 
is completed, press and hold “Set” (“Cal”) key for 5s or no operation within 10s to save data and exit 
automatically.

USB function module

Before  the  USB  function,  please  confirm  the  time  of  record  data  of  jump  drive  
interface.  If  the  time  does  not  match,  please  make  an  asdjustment  using  the 
aforementioned method.

Caution

Содержание HXC-106

Страница 1: ......

Страница 2: ...guidelines 5 Product Feature 6 Product Installation 7 Component Names Control Panel 12 Application method 16 Display and alarm 22 Cleaning and Maintenance 26 FAQ 28 Refrigeration Schematic Circuit Di...

Страница 3: ...en replacing the light from the power source to avoid electric hazards Warning 1 Unauthorized opening of the top cover of the control cabinet is prohibited to avoid damages to the parts inside or inju...

Страница 4: ...it is equipped with a 16A or 13A power plug Do not remove the ground pin of the power plug under any circumstances Make sure the plug is securely plugged into the power outlet to prevent re The power...

Страница 5: ...harmful or radioactive particles If the refrigerator is to be decommissioned unplug the power cord to avoid electric shock current leakage or re caused by aged power lines If the refrigerator is left...

Страница 6: ...accelerate the defrosting process other than those recommended by the manufacturer Do not damage the refrigerant circuit In order to reduce ammability HAZARDS the installation of this equipment shall...

Страница 7: ...e as the temperature in the center of the refrigerator The cabinet temperature will gradually reach an equilibrium state Only clean the unit with light soapy water Never use brushes acids gasoline soa...

Страница 8: ...lood bank refrigerator provided to you may be a little di erent from the gure in the manual due to product improvement The manual may be subject to changes without further notice 35 Name Blood Bank Re...

Страница 9: ...frigerator Do not install the refrigerator in a narrow and con ned space as heat may be trapped to increase ambient temperature Additionally a unit should not be installed in a space where the entry d...

Страница 10: ...14 1055 Interior dimensions W x D x H mm 460 370 950 430 350 830 E ective volume 158L 106L Door Glass door with electric heat Insulation CFC Free foamed in place urethane Compressor High Quality Herme...

Страница 11: ...mal operation 6 Installation of the Handle Handle installation of products shall be conducted in the following steps Orient the handle with the hole downwards Align the handle and handle guide on door...

Страница 12: ...ce the unit is powered up the audible alarm is turned on Turn on the battery switch and press the buzz cancel button to silence the alarm Visual alarm continues until the temperature in the measuremen...

Страница 13: ...oducts Do not load in too much warm product at one time Overloading the refrigerator will reduce the system s expected life Therefore product and samples should be placed into the refrigerator in batc...

Страница 14: ...g Temperature recorder Optional for HXC 158 Blood storage basket or drawer Shelf USB port optional Interior light Temperature measurement box Caster Door ajar alarm switch Air filter grille 29 Refrige...

Страница 15: ...perature is too high Air duct is blocked Unit seems to emit too much noise Check the installation of the unit to make sure the unit does sit on a solid oor without vibration Whether certain part of th...

Страница 16: ...temperature recorder is standard on the product optional for HXC 158 If additional chart paper is needed please contact Haier Biomedical The ordering cycle is 1 2 months Consult us by phone for detail...

Страница 17: ...e after sales service personnel Cleaning and Maintenance Warning Clean the refrigerator Replace the illumination light Condensate evaporator is located in the back of the refrigerator Remove the 2 scr...

Страница 18: ...tem is in a n alarm mode press the Buzzer Cancel key on the display board and the audible alarm is silenced temporarily Remote alarm condition will not be a ected After 30 minutes if the alarm conditi...

Страница 19: ...timely manner to protect the samples When there is a fault in the battery E5 will show after pressing Alarm test Check the battery charge switch to make sure it is in the On position If it is O turn...

Страница 20: ...ly Buzzing Alarm light ashes Display actual temperature Remote alarm contacts reverse and audible alarm starts Remote alarm can be cancelled after pressing Silence button If the alarm condition persis...

Страница 21: ...y battery is low ashing of LED indicator lamp on the recorder indicates that the battery needs to be changed Maintenance and change of emergency battery of the recorder When the green LED of recorder...

Страница 22: ...lay panel ashes USB which indicates the data are now exporting and not nished yet After 5 seconds the display will return to the temperature mode If the display shows ALL indicating the data export pr...

Отзывы: