Haier HXC-106 Скачать руководство пользователя страница 20

18

There are two automatic defrosting periods controlled by the controller.

1.Periodical Defrosting
The evaporator is embedded with electric heaters. During normal cycle off , the heaters are powered 
to  raise  the  evaporator  above  0

  to  thaw  out  the  ice. There  is  no  impact  to  the  refrigerator 

temperature during this time.

2.Forced Defrost
When  the  ambient  condition  is  severe  to  allow  excessive  moisture  enter  the  refrigerator,  the 
refrigerator will automatically initiate a forced defrost cycle. Once the defrost cycle is completed, the 
unit will resume  normal operation. During the  forced defrost time frame, it is expected to have the 
refrigerator temperature rise by about 1 to 2

.

Automatic Defrosting

Refrigerator temperature recorder of 6-inch standard panel is standard for our products(except HXC-
106/158) to record the temperature change in box. 
•Installation and Operation

In order to operate and use the recorder in a right way, please follow the steps below:
1. Open the door of recorder. You can see the recorder;
2. Switch on the 9V DC-battery on the right corner of the recorder. The battery is the emergency 
power supply for the recorder;
3. Examine the  recording  paper. If  it  needs to be  changed, please change the recording  paper 
according to the operation steps in P20;
4. Pull off  the plastic cap from recording pen and close the door of recorder

The recorder will not operate before the temperature in the box reaches the recording 
scope. 

Temperature recorder 

Caution

23

HXC-1308/1308B

SN.

Alarm items

Initiation of Alarm

Buzzing

Light

fl ashing

Display

requirements

Note

1

Over 

temperature 

alarm (high 

and low 

temperature 

alarm)

When the temperature of 

upper or lower temperature 

measurement box is 

warmer than the high alarm 

set point, the alarm starts. 

When the temperature 
of the upper and lower 

temperature measurement 

box is lower than the low 

temperature alarm set point, 

temperature alarm starts.

1Hz 

Continuously 

Buzzing

Alarm light 

fl ashes

Display actual 

temperature

Remote alarm contacts 

reverse and audible 

alarm starts. Remote 

alarm can be cancelled 
after pressing “Silence” 

button. If the alarm 

condition persists after 

30 minutes, the remote 

alarm and audible 

alarm resume.

2

Power failure 

alarm

When there is a loss of 
AC power to the unit or 

the power switch is in Off 

position, the power outage 

alarm will trigger. The 

rechargeable battery will 

power the display and alarm 

system.

1Hz 

Continuously 

Buzzing

The power 

indicator lights 

of alarm light 

fl ash at the 

same time

Alternatively display 
the upper and lower 

temperature in the 

measurement boxes 

for 5 seconds. The 

screen turns blank 

for 30 seconds, and 

resumes displaying the 

temperature.

Remote alarm contacts 

reverse and audible 

alarm starts. Remote 

alarm can be cancelled 
after pressing “Silence” 

button. If the alarm 

condition persists after 

30 minutes, the remote 

alarm and audible 

alarm resume.

3

Fault of sensor

Upper sensor is in short 

circuit or open circuit, fault 

code: E1

1Hz

Continuously

Buzzing

Alarm

light

fl ashes

Display board 

alternatively shows E1 

for 2 seconds and the 

temperature in the lower 

sensor for 6 seconds.

Remote alarm contacts 

reverse and audible 

alarm starts. Remote 

alarm can be cancelled 
after pressing “Silence” 

button. If the alarm 

condition persists after 

30 minutes, the remote 

alarm and audible 

alarm resume.

Lower sensor is in short 

circuitor open circuit, fault 

code: E2

Display board 

alternatively shows E2 

for 2 seconds and the 

temperature in the upper 

sensor for 6 seconds.

Control sensor is in short 

circuit or open circuit, fault 

code: E3

Display alternates 

showing E3 for 2 seconds 

and both temperatures 

in the upper and lower 

temperature boxes for 6 

seconds.

Defrosting sensor is in short 

circuit or open circuit, fault 

code: E4

Display alternates 

showing E4 for 2 seconds 

and both temperatures 

in the upper and lower 

temperature boxes for 6 

seconds.

Alarming sensor is in short 

circuit or open circuit, fault 

code: E5

Display alternates 

showing E5 for 2 seconds 

and both temperatures 

in the upper and lower 

temperature boxes for 6 

seconds.

Содержание HXC-106

Страница 1: ......

Страница 2: ...guidelines 5 Product Feature 6 Product Installation 7 Component Names Control Panel 12 Application method 16 Display and alarm 22 Cleaning and Maintenance 26 FAQ 28 Refrigeration Schematic Circuit Di...

Страница 3: ...en replacing the light from the power source to avoid electric hazards Warning 1 Unauthorized opening of the top cover of the control cabinet is prohibited to avoid damages to the parts inside or inju...

Страница 4: ...it is equipped with a 16A or 13A power plug Do not remove the ground pin of the power plug under any circumstances Make sure the plug is securely plugged into the power outlet to prevent re The power...

Страница 5: ...harmful or radioactive particles If the refrigerator is to be decommissioned unplug the power cord to avoid electric shock current leakage or re caused by aged power lines If the refrigerator is left...

Страница 6: ...accelerate the defrosting process other than those recommended by the manufacturer Do not damage the refrigerant circuit In order to reduce ammability HAZARDS the installation of this equipment shall...

Страница 7: ...e as the temperature in the center of the refrigerator The cabinet temperature will gradually reach an equilibrium state Only clean the unit with light soapy water Never use brushes acids gasoline soa...

Страница 8: ...lood bank refrigerator provided to you may be a little di erent from the gure in the manual due to product improvement The manual may be subject to changes without further notice 35 Name Blood Bank Re...

Страница 9: ...frigerator Do not install the refrigerator in a narrow and con ned space as heat may be trapped to increase ambient temperature Additionally a unit should not be installed in a space where the entry d...

Страница 10: ...14 1055 Interior dimensions W x D x H mm 460 370 950 430 350 830 E ective volume 158L 106L Door Glass door with electric heat Insulation CFC Free foamed in place urethane Compressor High Quality Herme...

Страница 11: ...mal operation 6 Installation of the Handle Handle installation of products shall be conducted in the following steps Orient the handle with the hole downwards Align the handle and handle guide on door...

Страница 12: ...ce the unit is powered up the audible alarm is turned on Turn on the battery switch and press the buzz cancel button to silence the alarm Visual alarm continues until the temperature in the measuremen...

Страница 13: ...oducts Do not load in too much warm product at one time Overloading the refrigerator will reduce the system s expected life Therefore product and samples should be placed into the refrigerator in batc...

Страница 14: ...g Temperature recorder Optional for HXC 158 Blood storage basket or drawer Shelf USB port optional Interior light Temperature measurement box Caster Door ajar alarm switch Air filter grille 29 Refrige...

Страница 15: ...perature is too high Air duct is blocked Unit seems to emit too much noise Check the installation of the unit to make sure the unit does sit on a solid oor without vibration Whether certain part of th...

Страница 16: ...temperature recorder is standard on the product optional for HXC 158 If additional chart paper is needed please contact Haier Biomedical The ordering cycle is 1 2 months Consult us by phone for detail...

Страница 17: ...e after sales service personnel Cleaning and Maintenance Warning Clean the refrigerator Replace the illumination light Condensate evaporator is located in the back of the refrigerator Remove the 2 scr...

Страница 18: ...tem is in a n alarm mode press the Buzzer Cancel key on the display board and the audible alarm is silenced temporarily Remote alarm condition will not be a ected After 30 minutes if the alarm conditi...

Страница 19: ...timely manner to protect the samples When there is a fault in the battery E5 will show after pressing Alarm test Check the battery charge switch to make sure it is in the On position If it is O turn...

Страница 20: ...ly Buzzing Alarm light ashes Display actual temperature Remote alarm contacts reverse and audible alarm starts Remote alarm can be cancelled after pressing Silence button If the alarm condition persis...

Страница 21: ...y battery is low ashing of LED indicator lamp on the recorder indicates that the battery needs to be changed Maintenance and change of emergency battery of the recorder When the green LED of recorder...

Страница 22: ...lay panel ashes USB which indicates the data are now exporting and not nished yet After 5 seconds the display will return to the temperature mode If the display shows ALL indicating the data export pr...