Maintaining the MCS+ Equipment
P/N 85213-30, Manual revision: B
Filter screens
The MCS+ device is equipped with filter screens on the bottom of the cabinet,
which eliminate dust from incoming cool air. The filters should be cleaned
routinely, especially if dust becomes visible on the screens.
To clean the filters, the operator should:
"
Remove the retainer plates using a Phillips-head screwdriver.
"
Remove the filter screens from the panel.
"
Rinse the screens under running water – DO NOT use any cleaning agents.
"
Gently squeeze the screens to remove excess water.
"
Place the screens on a clean, dry cloth and allow to dry completely.
"
Reinsert the screens into the panel, ensuring that all openings are com-
pletely covered by the filter.
"
Replace the retainer plates and tighten the screws.
Warning: To avoid electrical shock, the filter screen should be completely dry
before it is reinstalled on the MCS+ cabinet.
Bar-code reader
The bar-code reader window should be wiped using a lint-free cloth or gauze
and water, then dried. It should be cleaned whenever there is an accumulation
of dust or spilled fluid. For optimal cleaning, the operator should remove the
protector around the window and replace it once the window has been cleaned.
Pressure cuff
If it is determined that the cuff has been contaminated, clean the affected area
with a 10% bleach solution (or equivalent). If this cleaning is determined to be
insufficient based on the severity of the contamination, a replacement cuff may
be ordered and installed by the user.
Содержание LN9000-110-EWC
Страница 2: ...2002 2006 Haemonetics Corporation All rights reserved...
Страница 10: ...x Preface P N 85213 30 Manual revision B The Haemonetics MCS LN09000 220E ED...
Страница 20: ......
Страница 30: ......
Страница 81: ...Maintaining the MCS Equipment 5 7 P N 85213 30 Manual revision B...
Страница 85: ......
Страница 86: ...5 12 Maintaining the MCS Equipment P N 85213 30 Manual revision B...
Страница 100: ......