308
Operator Action Required
Explanation:
The bag line clamps must be set to continue Sequestration.
Corrective Action:
1. Set the bag line clamps as follows:
• CLOSE the white air line clamp.
• CLOSE the blue PRP line clamp.
• OPEN the yellow PPP line clamp.
2. Agitate the RBC bag.
3. Touch
Continue
to concentrate.
309
Blood Bag Is Empty
Explanation:
Air was detected while filling the bowl. Sequestration will continue using fluid from the RBC bag.
Corrective Action:
Touch
Continue
to continue.
310
Sequestration Cycle Complete
Explanation:
The sequestration cycle has completed.
Corrective Action:
Touch
Sequestration
to sequester another unit.
OR
Touch
Cell Salvage
to prepare for Cell Salvage.
OR
Touch
End Procedure
to end the current procedure.
Table 22, Event Messages