8. Close and latch the valve module cover.
Figure 51, Tubing harness
Installing the Bowl
To install the bowl:
1. Ensure that the lower port of the bowl faces the effluent line sensor.
2. Install the bowl in the centrifuge by carefully pressing down on the
shoulders of the bowl until it is seated securely in the chuck.
Figure 52, Inserting the bowl into the centrifuge chuck
3. Position the header arm around the top of the bowl.
4. Turn the latch on the header arm clockwise until it locks into place.
5. Spin the bowl to ensure it spins freely.
2.
3.
1.
4.
1. Tubing in air detector
2. Pump platen lever
3. Pump platen
4. Tubing in valve module
channels