180
Chapter 11
P/N 130859-US, Manual Revision: AA
Haemonetics
®
Cell Saver
®
Elite
®
+ User Manual
150
Device Cover Lock Failure
Explanation:
The device cover will not lock. The device cover must be closed and locked for the procedure to
continue.
Corrective Action:
1. Open and close the device cover.
2. Ensure that there are no obstructions to closing the cover.
If the problem persists, service is required.
151
Device Cover Open
Explanation:
The device cover is open. The device cover must be closed and locked for the procedure to
continue.
Corrective Action:
Close the device cover.
152
Fluid Detected in Centrifuge Well
Explanation:
The device detected fluid in the centrifuge well. There may have been a blood spill or an issue
with the fluid detector.
Corrective Action:
1. Check the effluent line for kinks and occlusions; ensure clamps are open.
2. Inspect the bowl and fluid sensor.
3. If a spill is observed, power down the device to clean the centrifuge per instructions in the
manual and replace the disposable.
4. If no spill is detected, once the centrifuge well has been dried, touch
Retry
.
If the problem persists, service is required.
Table 22, Event Messages
ID #
Event Message Text
Содержание Cell Saver Elite+
Страница 1: ...User Manual Not for use with software prior to revision AQ 130859 US AA October 2022...
Страница 4: ......
Страница 58: ......
Страница 72: ......
Страница 122: ......
Страница 154: ......
Страница 223: ...Haemonetics Cell Saver Elite User Manual P N 130859 US Manual Revision AA Appendix B System Performance Cell Salvage 224...
Страница 226: ......
Страница 229: ...Cart Assembly Instructions 229 Haemonetics Cell Saver Elite User Manual P N 130859 US Manual Revision AA...
Страница 230: ......