Ready to Remove Air from RBC Bag
Explanation:
To remove air from the RBC bag, the pump is started and stopped using the
Pump
pad.
Note:
Removal of air should stop prior to RBCs leaving the RBC bag.
WARNING:
This process may leave residual air in the RBC bag. Do not pressure infuse. May
cause fatal infusion of air.
Corrective Action:
1. Hold the RBC bag with the blue line facing up.
2. Touch and hold
Pump
to remove air from the reinfusion bag.
3. Release
Pump
to stop the pump.
283
USB Flash Drive Error
Explanation:
An error occurred while transferring data with the USB flash drive. The operation did not
complete successfully but can be retried.
Corrective Action:
1. Verify the flash drive is inserted properly.
2. Touch
Continue
to continue.
3. Retry the operation.
If the problem persists, a different USB flash drive or service may be required.
284
Data Transfer in Progress
Explanation:
The requested data transfer operation is in progress.
Corrective Action:
Wait for the operation to complete before removing the USB flash drive.
Table 22, Event Messages
ID #
Event Message Text
Содержание Cell Saver Elite+
Страница 1: ...User Manual Not for use with software prior to revision AQ 130859 US AA October 2022...
Страница 4: ......
Страница 58: ......
Страница 72: ......
Страница 122: ......
Страница 154: ......
Страница 223: ...Haemonetics Cell Saver Elite User Manual P N 130859 US Manual Revision AA Appendix B System Performance Cell Salvage 224...
Страница 226: ......
Страница 229: ...Cart Assembly Instructions 229 Haemonetics Cell Saver Elite User Manual P N 130859 US Manual Revision AA...
Страница 230: ......