4.4. Changing the battery
The charging state of the battery is monitored by the portable strain gauge force gauge. If the battery charge is
lower than 4V, “Batt” flashes in the bottom line of the display. In this case, for safe operation of the force gauge,
the battery needs to be replaced.
To replace the battery, the PAD2 must be switched off. Please keep four new batteries type AA at hand.
Remove the blue protective cover from the housing of the PAD2. The battery compartment is at the rear of the
device.
With slight pressure on the ribbed surface, slide the lid of the battery compartment downwards beyond the
housing.
Remove the used-up batteries and insert the new ones. Observe the batteries‘ polarity in doing so.
Slide the lid along the lateral guide back onto the battery compartment and let the nose click into place in the
housing by applying slight pressure.
Thread the housing first on the cable side and then on the opposite side back into the protective cover.
4.5. Turning off:
By holding the button down, the force gauge is switched off.
The peak values are deleted. If the system was tared, this tare value is kept even if the PAD2 is switched on or
off. The individual settings entered in the menu are kept.
5. Calibration cycle
The operational reliability and accuracy of the PAD2 and any existing test weights must be checked at regular
intervals according to the quality assurance regulations.
For those responsible this force gauge need to determine a suitable interval in which it is regularly presented to
HAEHNE
Elektronische Messgeräte GmbH for testing. The type and scope of this test also need to be stated
by those responsible.
We recommend test intervals of one year or two years for checking and recalibrating the operational reliability
of the PAD2.