Haehne PAD2 Скачать руководство пользователя страница 7

Page 7    

PAD2 EA EN 08_16.indd

HAEHNE

  Elektronische Messgeräte GmbH · Heinrich-Hertz-Str. 29 · D-40699 Erkrath Germany · Telefon 0211/9 25 91-0 · Fax 0211/9 25 91-20

http://www.haehne.de 

Email: [email protected]

Contrast

Here you can set the contrast in the display, to be most readable for you.
You can select between 10 - 100 % in 10%-steps

Procedure:

Press 

MENU

 to open the menu

Select the “2. Configuration” menu option with 

 and open it with 

enter

Select the desired submenu option with 

 and open it with enter

Select the required value with 

 and confirm it with 

enter

Select another submenu option with 

 

or
Select the submenu option “2.4 Back” with 

 and return to the main menu with 

enter

or

Exit the menu with 

escape

. The device returns to measuring mode.

Note

If, while entering the values, you exit the menu with 

escape

 without 

having confirmed the entered values with 

enter

, then the entered 

value is not saved.

4.3. Measuring
Warning

There is risk to life and limb of the operator and other persons involved in case of incorrect operation or misuse  
of the portable strain gauge force gauge PAD2 or non-observance of the safety regulations. The force gauge or 
other  
material assets may be damaged.

To adapt the PAD2 to existing lifting devices, only lifting gear (swivel heads, eyebolts etc.) may be used that 

 

correspond at least to the nominal force* of the sensor. The operator is responsible for selection and control.  

Product liability and warranty of the manufacturer exclusively refers to the standard equipment of the force gauge 

 

and the fixtures supplied ex factory.

The PAD2 must not adapted under load.
Before using the force gauge, the operator needs to check all load suspension devices (swivel heads, eyebolts 
etc.) for secure fit and damages (fissures, cracks, deflections etc.). Damaged devices must no longer be used.

When measuring, the lifting of a load, e.g. with a lifting device, must be slow and steady. Fast or unsteady load 

pick up results in dynamic load peaks which can lead to overload, damages up to breaking of the load  
pick up devices or measuring device. In general, the applied load, consisting of d tared load, must 

not exceed the nominal force* of the sensor!

Getting under a suspended load or entering the direct risk area is prohibited!

* Information about the nominal force of the sensor can be found on its type label or you can read the nominal 
force in the display when switching on the force gauge.

Содержание PAD2

Страница 1: ... Description 2 1 2 Scope of supply 2 1 3 Intended use 2 1 4 Safety instructions 2 2 Technical Data 3 3 Operating Elements and Functions 4 4 Operation 5 4 1 Switching on 5 4 2 Using the menu MENU 5 4 2 1 About menu option 6 4 2 2 Configuration menu option 6 4 3 Measuring 7 4 3 1 Zero adjustment and taring 0 T 8 4 3 2 Switching the measured value display PEAK 8 4 3 3 Performing the measurement 8 4 4...

Страница 2: ... its nominal load The nominal load can be found on the type label of the sensor You can also find this infor mation on the force gauge display when switching it on The force gauge must not be subjected to any forces This force gauge is intended only for the exclusive use in areas that are not subject to calibration requirements It is a highly sensitive measuring instrument and must be treated acco...

Страница 3: ...orce gauges display when switching on Display 12 digits 2 lines Resolution 16 bit Power supply of PAD2 4 x battery type AA Operating time 8 h continuous operation Power supply of force sensor 3 3V Nominal resistance of strain gauge 100Ω to 500kΩ Auto off time 8 30 60 min off Operating temperature range From 0 to 50 C 32 to 122 F Storage temperature range From 20 to 70 C 4 to 158 F Protection class...

Страница 4: ...n this mode the force gauge is calibrated in the factory Button Name in the measuring mode Function in the measuring mode grey font Name in the menu Function in the menu light blue font ZERO POSITIONING TARING Zero adjustment taring and Clear peak values up scroll up in the menu increase numeric values PEAK Toggle between displays of the actual measured value of the max value largest value in posi...

Страница 5: ...ake sure that this load is not exceed ed With the connected sensor the current measured value is shown in the display Please note that a previously tared value has been retained The peak values were deleted 4 2 Using the menu MENU The menu provides the following options Read out information about the device About menu option Configure device settings Configuration menu option Calibrate the device ...

Страница 6: ...be read in the display Return to the submenu with enter Select another submenu option with or Select the submenu option 1 4 Back with and return to the main menu with enter or Exit the menu with escape The device returns to measuring mode 4 2 2 Configuration menu option With this menu option you can configure the following settings Auto off Select the time after which the device automatically swit...

Страница 7: ...afety regulations The force gauge or other material assets may be damaged To adapt the PAD2 to existing lifting devices only lifting gear swivel heads eyebolts etc may be used that correspond at least to the nominal force of the sensor The operator is responsible for selection and control Product liability and warranty of the manufacturer exclusively refers to the standard equipment of the force g...

Страница 8: ...xiting the menu the current measured value is always shown in the display You can also select display of the peak values with the PEAK button In this setting the measured value display corresponds to a drag indicator which is further advanced in case of increasing decreasing values If you press the PEAK button once then the maximum value is displayed in positive load direction Peak Max is displaye...

Страница 9: ...teries polarity in doing so Slide the lid along the lateral guide back onto the battery compartment and let the nose click into place in the housing by applying slight pressure Thread the housing first on the cable side and then on the opposite side back into the protective cover 4 5 Turning off By holding the button down the force gauge is switched off The peak values are deleted If the system wa...

Страница 10: ...NE Elektronische Messgeräte GmbH Heinrich Hertz Str 29 D 40699 Erkrath Germany Telefon 0211 9 2591 0 Fax 0211 9 2591 20 http www haehne de Email info haehne de PAD2 Technical modifications reserved Page 10 ...

Страница 11: ... auto off averaging measurement rate setting and Contrast These are described in the Operating Manual PAD2 The third menu item contains the factory settings calibration This area is password protected with a 4 digit number During each calibration a counter is counted up irrevocably this count reading is displayed in each case before the re calibration The settings in the service menu 3 must be con...

Страница 12: ...ounter If canceled with Esc the counter is not increased Note Termination is currently available only at the display Load is still changed Load Here the applied test load can be specified in 10 increments of the rated load Pkt 1 min The zero point is set without load Pkt 2 max Max is set with X of load The calibration process is completed successfully for 1 5 seconds Cal Ready appears on the displ...

Страница 13: ... level 3 Residential area Interference field strength Industrial sector Electrostatic discharges ESD Electromagnetic fields Electromagnetic fields mobile phone radiation Before commissioning the PAD2 it is necessary to ensure that the system consisting of the portable strain gauge force gauge PAD2 and a connected sensor is according to the EC Guidelines EMC in case of modifications to the instrume...

Страница 14: ...HAEHNE Elektronische Messgeräte GmbH Heinrich Hertz Str 29 D 40699 Erkrath Germany Telefon 0211 92591 0 Fax 0211 92591 20 http www haehne de Email info haehne de HAEHNE ...

Отзывы: