3. Operating elements and functions
The portable force gauge has two operating modes:
- the measuring mode:
in this mode, the measurements are performed. This is the standard mode of the PAD2, which is always enabled
after switching on.
- the menu mode:
The most important data about the force gauge can be read out and configured.
In this mode, the force gauge is calibrated in the factory.
Button Name in the
measuring
mode
Function in the measuring
mode
(grey font)
Name in
the menu
Function in the menu
(light blue font)
ZERO
POSITIONING/
TARING
- Zero adjustment/taring and
- Clear peak values
up
-
scroll up in the menu
-
increase numeric values
PEAK
Toggle between displays:
-
of the actual measured value
-
of the max. value (= largest
value in positive force direction)
-
of the minimum value (= largest
absolute value in negative force
direction)
down
-
scroll down in the menu
-
decrease numeric values
ON/OFF
Switch on/off
escape
-
exit sub menu option during
entry without saving
-
exit menu, return to measuring
mode
MENU
Open the menu
enter
-
open selected menu option
-
confirm and save entry