The menu is used via the four buttons.
- It is opened with the
MENU
button.
- You can scroll through the menu options / submenu options with the
buttons, or increase or decrease
adjustable values.
- The
enter
button is for opening the selected menu option or to confirm the entry.
- The menu is closed with the
escape
button. The device is then in measuring mode.
4.2.1. „About“ menu option
With this menu option, you can read out the following information about your device:
Firmware
Version of the installed software
Data last cal. Date of the last calibration
Serial No.
Serial number
Procedure
:
Press
MENU
to open the menu
Open menu item “1. About” with
enter
Select the desired submenu option with
and open it with
enter
The device-specific information can be read in the display
Return to the submenu with
enter
Select another submenu option with
or
Select the submenu option “1.4 Back” with
and return to the main menu with
enter
or
Exit the menu with
escape
. The device returns to measuring mode.
4.2.2. „Configuration“ menu option
With this menu option, you can configure the following settings:
Auto off
Select the time after which the device automatically switches off to save batteries.
You can select one of the following: off, 60, 30, 8 min. (Factory setting: 60 min)
Averaging
Here, you can select the desired measuring rate customized for your application, at the
same time, the frequency also changes with which the measured value display is up-
dated in the display. The lower measuring rate is suited to slower loads, the higher rate
is suited to short, fast load changes. The higher the measuring rate, the better the detec-
tion of peak loads. However, with the high measuring rate the measured value display
becomes slightly more disturbed, as the noise element increases.
You can select one of the following:
Settings
Measuring
rate
Display frequency
slow
10 Hz
2 Hz
middle
20 Hz
5 Hz (factory setting)
fast
100 Hz
10 Hz