Safety Instructions
chuck (test this with a dipstick and clean if
necessary),
•
the inserted tool sits properly in the chuck
and will not cant/jam while clamping it,
•
the tool is approved for using it at the max-
imum speed (see manufacturer’s instru
c-
tions) or only use the speed required by the
instrument,
•
the tool is absolutely dry.
For safety reasons do not change the tool
while the motor is running.
Switch off the unit immediately if
•
the bur comes off during the treatment it-
self.
•
the motor blocks during the treatment.
While doing any foot care, only operate the
unit with the suction turned on.
Only use the suction function, if you have in-
stalled the suction filter. Check this filter every
10 operating hours. If it is full, please ex-
change it. A filter should be exchanged at
least after 50 operating hours.
Never use the suction to vacuum any liquids
(or air/water fog which contains particles of
liquids).