16
Application
As an option you may also use the
micro filter (art. 5179) instead of
paper filter (art. 5115).
After the filter bag has been
changed, the operating time must
be set to zero. To do this, set the
device to standby by pressing but-
ton (9). Now hold down button (2)
for five seconds until
FC
appears in
the display.
Exchange Control Unit Filter
This filter (art. 3752) has to be ex-
changed after every 200 operating
hours. Remove the old filter and
place the new one in its holder (see
picture).
Clamping the Bur/Tool
The handpiece is equipped with a
chuck, which can hold all burs with
a diameter of 2.35 mm (this is the
diameter usually used for tools in
podiatry). To open the chuck, push
the knob forward with your thumb
and leave it there. Now insert the
tool as far as possible. Pull back the
knob and the tool will be firmly
clamped.
For security reasons, only
change the tool, if the
handpiece is turned off.
Setting up the Speed
Hold the handpiece (incl. tool) in
your hand or store it in the hand-
piece holder attached to the side of
the control unit. Switch on the unit
with button (9). Now you can also
set the speed with buttons (1) and
(2) and choose its direction of rota-
tion with button (6).
When the device is brought
out of standby mode it will
run with the rotation speed
and suction power used last. Take
care to not start the mill at an overly
high rotation speed. Select the re-
quired rotation speed before chang-
ing the milling tool.
Foot Switch (art. 0970)
Operation of the foot switch
Only use the original hadewe foot
switch, to avoid any damages. In-
sert the plug of the foot switch into
the socket with the symbol
.
If you press the foot switch for a
longer period, you will switch on/off
the complete control unit.