background image

 

18 

Usage 

and  (4).  Choose  the  direction  with 
icon (18). 
 
 

Memory Keys  

Frequently  used  speed  can  be 
saved. There are 3 memory spaces 
available.  First  set  the  desired 
speed. Save this by pressing one of 
the three memory keys (15), (16) or 
(17)  for  4  seconds.  The  saved 
speed  is  now  directly  available  on 
pressing the memory key. To recall 
the  stored  speed  quickly,  briefly 
press  the  corresponding  memory 
key (15), (16) or (17). 
 

 

Only  use  burs/tools  which 
are  permitted  for  speed 
level  saved  under  (15)  to 
(17). 
 

 

When  the  device  is  brought 
out  of  standby  mode,  it  will 
run  with  the  rotation  speed 
and  suction  power  used 
last.  Take  care  to  not  start 
the  mill  at  an  overly  high 
rotation  speed.  Select  the 
required 

rotation 

speed 

before  changing  the  milling 
tool. 

Foot Switch  

Operation of the foot switch 

Only  use  the  original  hadewe  foot 
switch  (art.  no.  0970)  to  avoid  any 
damages. Insert the plug of the foot 

switch  into  the  socket  with  the 
symbol 

Function of foot switch 

By holding down the foot switch, the 
power  can  be  switched  ON/OFF. 
While the spray is on, and you push 
the  switch  once,  the  spray  stream 
will  stop.  If  you  push  it  a  second 
time,  the  handpiece  will  stop. 
Pushing  the  switch,  a  third  time, 
both  functions  will  start  to  work 
again. 
 
By  briefly  pressing  the  foot  switch, 
the  handpiece  and  spray  are 
stopped  (e.g.  to  change  the  tool/ 
bur).  By  pressing  again  briefly,  the 
handpiece  and  spray  can  be 
reactivated.   
 

 

When 

the 

display 

is 

switched  off,  you can turn  it 
on  by  keeping  the  foot 
switch 

pressed 

(keep 

pressed long). 

 
 

Содержание Azuras CHROME

Страница 1: ...art 0982 FB 04 303 Rev 1 Date 30 01 2019 Manual Azuras CHROME ...

Страница 2: ...ay Speed 17 Spray 17 Insert Bur 17 Adjusting the Speed 17 Memory Keys 18 Foot Switch 18 Select Menu 19 Help Menu 19 Settings Menu 20 Confirm Washer Change 20 Turn Light ON OFF 20 Pictures Videos 21 Messages 24 Maintenance Care 25 Washer 25 Cleaning the Chuck 25 Water Filter 26 Cleaning Function 26 Maintenance Overview 27 Upload Firmware 27 Handling Instructions 28 Troubleshooting 29 Technical Deta...

Страница 3: ...ice for the treatment of callus and nails The handpiece is compatible with rotating tools such as diamond polishers steel cutters and ceramic cutters These can be worked with up to a speed of 40 000 rpm thus providing an efficient way of removing tissue The direction of rotation can be changed by pressing a button The integrated spray function emerges reliably bundles emerging dusts The rotation s...

Страница 4: ...e if they have been given supervision or instruction concerning use of the appliance in a safe way and understand the hazards involved Children should be supervised not to play with the appliance Cleaning and user maintenance shall not be made by children without supervision To avoid the risk of electrocution only connect the device to the main power supply If you pull the power plug it will stop ...

Страница 5: ...h have it changed or repaired by an authorized service center Prevent your hair from wrapping itself around moving parts If applicable wear a hair net When working with materials which might create dust or moisture use a drilling unit with suction or spray technique Wear a nose and mouth protection Always follow the safety instructions provided by the manufacturer of the materials being formed and...

Страница 6: ... additive e g hadewe AquaHy art no 8555 If you use other brands pay attention to their manuals if you can use them for this purpose Never use tap water or alcoholic fluid or other burnable liquids In case you are working with an alcoholic fluid or other burnable liquids there is always the danger of fire or explosion hazards Never refill the tank directly over the unit to avoid in case you acciden...

Страница 7: ...firmly in the chuck do not work with tools which have an oily shaft are worn are bent Before using the equipment assure that the tool can be pushed far enough into the chuck test this with a dipstick and clean if necessary the inserted tool sits properly in the chuck and will not cant jam while clamping it the tool is approved for using it at the max speed see manufacturer s instructions or only u...

Страница 8: ...pecialist The mains cable may also only be changed by the manufacturer or an authorized specialist Maintenance records may be supplied to repair workshops and trained technicians upon request Never remove the housing especially not while the equipment is plugged into the socket since some components are under electric voltage Any unauthorized or improper handling e g trying to repair the unit of t...

Страница 9: ... used last Take care to not start the mill at an overly high rotation speed Select the required rotation speed before changing the milling tool In case the touch display shows a damage or does not work properly the device may not be used any longer Please send it to a qualified service workshop for repair ...

Страница 10: ...After every treatment please carry out the following steps Remove the rotating tool e g bur capper etc first Then handle the tool according to the instructions of its manufacturer Remove tissue remains by wiping off the unit with a mildly moistened paper cloth Afterwards clean all surfaces with disinfection cloth e g Mikrozid AF Tücher of Schülke and Mayr There is no limit to the number of times t...

Страница 11: ...t off standby 6 spray regulator knob 7 time 8 date 9 speed display 10 turn OFF ON spray 11 maintenance required 12 USB stick plugged in green light 13 light yellow when switched on 14 turn off handpiece spray turn on airflow 15 17 memory keys for speed 18 changing the direction of rotation left right switch 19 menu with additional functions 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 ...

Страница 12: ... following scopes foot care milling trimming and polishing manicure grinding milling and polishing works The devise is determined of operating rotating instruments which will not penetrate the skin or murcous membrane if used appropriate but could evoke damages accidently Caution In the area of pedicure you have to work with spray techniques Operation The working area must conform to the condition...

Страница 13: ...ill find the main power switch Press the main power switch 1 to turn the unit on The display prompts you to set the device language This is followed by 5 other pages with important clues With continue you proceed to the next page On the last page press to start the device The counter for calculating the maintenance proposal will automatically be set to zero To use the handpiece and spray function ...

Страница 14: ...ay be useful if the resulting milling dust is swirled in a clockwise direction in the field of view Now you can turn on the spray by pushing button 10 By using the spray regulator knob 6 you can set the amount of the spray liquid After setting the speed and the spray amount you can start with working Frequently used speed can be saved First set the desired speed Save this by pressing one of the th...

Страница 15: ...15 Usage Work Positioning The following images are examples of how to work on various parts of the foot Always make sure that the foot is firmly gripped and that you can work freely and without strain ...

Страница 16: ...16 Usage ...

Страница 17: ... Please only use distilled water together with disinfectant AquaHy Other admixtures could be unsuitable and cause damages at your device Particularly alcohol based spray fluids often attack hoses and make them porous Spray Turn on the spray function while pushing switch 10 Now the spray system will be filled automatically and a strong spray stream will be blowed out of the handpiece for the first ...

Страница 18: ...gh rotation speed Select the required rotation speed before changing the milling tool Foot Switch Operation of the foot switch Only use the original hadewe foot switch art no 0970 to avoid any damages Insert the plug of the foot switch into the socket with the symbol Function of foot switch By holding down the foot switch the power can be switched ON OFF While the spray is on and you push the swit...

Страница 19: ... date and language With button you can return to the previous menu Help Menu Manual Here is a short form of the manual This does not replace the original manual in paper form as the latter contains more detailed instructions and safety instructions Insert bur this clip demonstrates the clamping of rotary tools Change washer video clip that shows how to change the filter bag Change filter video cli...

Страница 20: ... approved for a speed of 40 000rpm the safety message can be deactivated Demo video activate the demo video Confirm Washer Change Every 50 hours of operation you will get the message to Please change washer After the washer change the following must be entered into the instrument On the start screen click the icon for the select menu Then click Change washer Confirm the filter change with YES Turn...

Страница 21: ... 100 files may be located in the directory for the image viewer Click Pictures Videos in the select menu to start the viewer file list scrollbar thumbnail reset In the file list you can see the files from the USB stick By scrolling down you can browse through the list Select an image and a small preview is shown on the right Click on thumbnail to enlarge it Click once more on the enlarged view of ...

Страница 22: ...eo library we provide a program ready to convert your own videos from other formats to EMF format Open the file to begin installation Follow the instructions on the monitor The conversion program to be used is File Converter Input Select a source file The following formats are supported MP4 WMV The source files should be proportionate to 4 3 to be taken so as to avoid distortion when converting Ou...

Страница 23: ...rive in order to consider the devices In the video library you can find videos and pictures that help you to explain the treatment to your customers Especially new customers are often unsecure because they do not know what to expect during a treatment It is helpful to explain the treatment steps to the patient This is often difficult especially when the patient cannot see the treatment site Pictur...

Страница 24: ...ol will appear Tapping this icon will display the following message Press OK to close this message In this case contact your distributor or repair shop CHANGE WASHER This message will come after 50 hours of operating Please change the washer See this manual page 25 SHUTDOWN DUE TO SPEED DIFFERENCE The device switches off when the set handpiece speed cannot be maintained SHUTDOWN DUE TO HIGH LOAD D...

Страница 25: ...nternal parts You will receive a message in case it has to be changed Exchanging the washer Remove the washer with tool art no 6475 Put on the tool between the LEDs If you set the tool on the LED they may break Look for little parts of the washer in the top and around If you find some remove such eventually Push the new washer in the top Confirm the washer change see page 20 Cleaning the Chuck Cle...

Страница 26: ...rst For its instructions see paragraph above Fill the tank with distilled water Close the tank and put it back into the unit Adjust the amount of spray regulator knob 6 on maximum Now press icon Clean in the display under Select Menu The complete cleaning process will last three minutes and then switches off by itself If requested you could cancel the process by pushing button 5 During the cleanin...

Страница 27: ... www hadewe net video library Save the file on a USB stick The file must be located in the root directory The file name encrypted_blto_m7 bin must not be changed 2 Turn on the device with the main power switch 1 and plug in a USB stick 3 Simultaneously press the buttons and hold down while you switch on the device using the main power switch 1 4 Display during the update This process takes approxi...

Страница 28: ...well known manufacturers formation of rust can appear e g through wrong treatment with chemical cleaning and disinfection liquids This extraneous rust affects internal parts of the handpiece in form of rust bloom and causes pitting Therefore when you not working with the handpiece always remove the bur tool from it Always make sure that you do not let the handpiece fall down or exposes it to any s...

Страница 29: ...cause Pull out the mains plug to reset the device After lengthy use unit does not spray Unit is overheated Slots of the unit should not be covered or blocked Let cool down the unit for a while The unit shows different sound levels at different sites of use The sound level varies according to the surroundings of the unit If you set up the unit directly in front of a wall the sound is reflected stro...

Страница 30: ...r 6354 USB stick 4 GB 6475 tool for washer 8555 AquaHy The clamping system is designed for rotating instruments which have a shaft of 2 35 mm in accordance with DIN EN ISO 1797 1 Operating Time The device is designed for conti nuous operation Ambient Conditions The device is intended for use in dry closed rooms Operation Temp 10 C to 26 C Rel humidity 25 to 75 not condensed Air pressure 700 hPa to...

Страница 31: ...under our sole responsibility that the product podiatry unit 0982 to which this declaration relates are conform to the following standards or normative documents 2014 30 EU directive EMC 2014 35 EU low voltage directive hadewe GmbH Grambartstraße 10 30165 Hannover Germany Safety Check A safety check must be done in compliance with the regulations of the country where the device is used The inspect...

Страница 32: ...hadewe GmbH Grambartstraße 10 30165 Hannover Germany info hadewe de www hadewe de ...

Отзывы: