Usage
Infilling the Spray Liquid
Switch off the device. Pull out the
spray tank on its handle. Hold the
cover of the tank in hand and
unscrew the tank. Now fill it with
distilled water and anti-germ liquid,
e.g. AquaHy (art. no. 8555).
Do not exceed the maximum
level when filling the tank.
Now turn the tank cover back on the
tank. While screwing back the cover
on the tank, make sure that is
closed correct. Otherwise there will
no pressure be built up. Put the
tank back in the housing. Switch on
the power switch in the rear of the
unit.
Attention: Never use tap
water (danger of calcification).
Also, do not use explosive or
combustible
liquids,
e.g.
alcohol.
Note: Never refill the tank
directly over the unit to avoid
in case you accidently spill
water, this will flow into the
unit.
The water filter (art. no. 3479)
has to be fully under water,
otherwise no water will be
sucked in.
Please only use distilled
water
together
with
disinfectant AquaHy. Other
admixtures
could
be
unsuitable
and
cause
damages at your device.
Particularly
alcohol-based
spray fluids often attack
hoses
and
make
them
porous.
Spray
Turn on the spray function while
pushing switch (10). Now the spray
system will be filled automatically
and a strong spray stream will be
blowed out of the handpiece for the
first 7 sec. Then you can adjust the
requested spray stream while
turning knob (6).
When turning off the unit or the
spray function, the spray liquid is
pumped back completely into the
tank from the hoses. With this
process, a bubbling can be heard in
the tank.
Insert Bur
Insert the bur until it stops. The
handpiece is equipped with a chuck
which can hold all burs with a
diameter of 2.35 mm (confirm to
DIN EN ISO 1797-1).
For security reasons, only
change the tool if the
handpiece is turned off.
Adjusting the Speed
Put the handpiece back in the
handpiece holder or keep it in your
hand. Switch ON the unit with
switch (5). Now you can adjust the
requested speed with the switch (3)