Table 22 Electrode
Parameter
Description
Basic mode Expert mode
Type
Type of measurement electrode
Part of name
Part of the name used to identify matching electrodes if the
preferred electrode is not connected
Recommended electrode
Recommended electrode to be used for this analysis
x
Calibration frequency
Recommended frequency of calibration (0 = no reminder)
x
x
Stability criterion
Refer to the working procedure for the recommended value
x
Max. stability time
Maximum time limit for the stability check
x
Stirring speed
Speed of the stirrer during electrode calibration
x
x
pH electrode parameters
Stirring duration
Stirring duration before buffer measurement (0 = buffer
measurement with stirring)
x
Max. temp. variation
Maximum permitted temperature variation during calibration
Min. slope
Minimum permitted slope limit for the calibration result
x
Max. slope
Maximum permitted slope limit for the calibration result
x
Min. offset
Minimum permitted offset limit for the calibration result
x
Max. offset
Maximum permitted offset limit for the calibration result
x
Calibration mode
Set the electrode calibration mode
x
Buffer set
List of available buffers for electrode calibration
x
x
Number fixed buffers
The number of fixed buffers to use
Fixed buffer #n
Buffer to be used during calibration
x
Conductivity electrode parameters
Calibration mode
Set the electrode calibration mode
x
Standard
Standard used for electrode calibration
x
x
Min. cell constant
Minimum permitted cell constant for the calibration result
x
Max. cell constant
Maximum permitted cell constant for the calibration result
x
Temp. compensation
Follow Natural Water Standards ISO/DIS_7888
x
Linear temp. coeff.
Coefficient for linear temperature compensation mode
x
Reference temp.
Reference temperature for linear temperature compensation mode
x
PtPt electrode parameters
Working mode
Specify if the probe is used in imposed voltage/current (AC or DC)
x
Current (sample analysis) Set the value of the imposed current during sample analysis
x
Current (titrant calibration) Set the value of the imposed current during titrant calibration
x
Voltage (sample analysis) Set the value of the imposed voltage during sample analysis
x
Voltage (titrant calibration) Set the value of the imposed voltage during titrant calibration
x
Change application parameters
64
Содержание TitraLab AT1102
Страница 1: ...DOC022 52 93074 TitraLab AT1000 series workstations User Manual 03 2019 Edition 7...
Страница 2: ......
Страница 6: ...Specifications 4...
Страница 20: ...Installation 18...
Страница 25: ...Install accessories 23...
Страница 26: ...Install accessories 24...
Страница 34: ...Startup 32...
Страница 42: ...Standard operations 40...
Страница 50: ...Maintenance 48...
Страница 58: ...Troubleshooting 56...
Страница 64: ...Replacement parts and accessories 62...
Страница 74: ...Change application parameters 72...
Страница 79: ......