3.7 Connect the tubes
Arrow symbols identify the inlet and outlet ports for the syringe and the pump
connections. The “up” arrow is the outlet port. The “down” arrow is the inlet port. Turn the
tube connectors on the inlet and outlet ports of the syringe and pump until they click.
The syringe outlet tube has a blue ring on it. If anti-diffusion tips are necessary, remove
the pre-installed outlet tube from the syringe and install the tube from the application kit
with the pre-installed anti-diffusion tip.
Push the outlet tubes into the tube holder slots so that they are correctly attached.
3.8 Install the sensor
3.8.1 Install the legacy adapter
Go to
on page 15 if no legacy adapter is included in the application
kit.
1.
Connect the measuring, reference and temperature sensors to the legacy adapter.
2.
Make sure that the instrument display shows the Home screen. Connect the legacy
adapter cable to a sensor socket on the rear panel of the instrument.
Installation
14