Section 7
Standard operations
C A U T I O N
Multiple hazards. Only qualified personnel must conduct the tasks described in this
section of the document.
C A U T I O N
Chemical exposure hazard. Obey laboratory safety procedures and wear all of the
personal protective equipment appropriate to the chemicals that are handled. Refer to the
current safety data sheets (MSDS/SDS) for safety protocols.
C A U T I O N
Personal injury hazard. Never use the instrument without the syringe protection cover in place.
C A U T I O N
Chemical exposure hazard. Never remove the stir bar from the beaker before the end of a titration.
7.1 Calibration
7.1.1 Calibrate the sensor
1.
From the home screen, select
Calibration
, then push
Electrode calibration
.
2.
If more than one sensor is installed, push the up and down arrow keys to highlight the
sensor to use, then push
Select
.
3.
If more than one application includes calibration parameters for the sensor, push the
up and down arrow keys to highlight the application to use, then push
Select
.
Calibration information shows on the display.
4.
If necessary, select an icon for more information or to change some data.
Option
Description
Electrode
Shows more information about the sensor.
Operator
Changes the operator ID. Select from a list of applicable operators.
Buffer or Standard
Shows more information about the buffer set or standard.
5.
Do the instructions that show on the display, then push
Start
to start the calibration.
Calibration data shows on the display.
6.
If the default stirring speed needs to be adjusted, push the up and down arrow keys
to increase or decrease the speed.
Note: This adjustment only applies to the current operation. The standard default stirring speed
for the calibration is not changed.
7.
Select
Stop
at any time to stop the calibration. Results are then calculated from the
calibration data available before
Stop
is selected.
8.
For pH sensors only:
Option Description
Yes
Continue with the next calibration buffer solution in the sequence.
No
Stop the calibration. The calibration can still be validated if at least one buffer
calibration was successful.
33
Содержание TitraLab AT1102
Страница 1: ...DOC022 52 93074 TitraLab AT1000 series workstations User Manual 03 2019 Edition 7...
Страница 2: ......
Страница 6: ...Specifications 4...
Страница 20: ...Installation 18...
Страница 25: ...Install accessories 23...
Страница 26: ...Install accessories 24...
Страница 34: ...Startup 32...
Страница 42: ...Standard operations 40...
Страница 50: ...Maintenance 48...
Страница 58: ...Troubleshooting 56...
Страница 64: ...Replacement parts and accessories 62...
Страница 74: ...Change application parameters 72...
Страница 79: ......