
29
SECTION 3,
continued
10. To save the calibration and return to the Reading mode, press
ENTER
. To exit the calibration without saving it and return to
the reading mode, press
EXIT
.
3.3
Reviewing the Calibration
1. From the pH Reading mode, press the
REVIEW
key.
2. The display will show the standard number, standard pH, and
temperature (view mV value by pressing
pH/mV
). Press the up
arrow once.
3. The meter will continue to scroll through the standard
information with each press of the up arrow key. When all the
standards have been displayed, press the up arrow key again.
4. The meter will display the slope and offset of the
calibration curve.
5. To exit Cal Review mode, press
EXIT
. To review any standard
or slope information again, press the down arrow.
Содержание sension1
Страница 7: ...7 SAFETY PRECAUTIONS Section 2 2 on page 19 Battery Installation ...
Страница 8: ...8 ...
Страница 10: ...10 ...
Страница 12: ...12 ...
Страница 14: ...14 SECTION 1 continued Figure 1 Hach sension1 Meter Figure 2 Electrode Holder ...
Страница 24: ...24 ...
Страница 34: ...34 ...
Страница 36: ...36 ...
Страница 38: ...38 REPAIR SERVICE ...