DISSOLVED OXYGEN PROBE, 

continued

General Probe Operation

Follow the procedures presented in this section to obtain maximum 
performance and accuracy from your DO probe:

Use the DO probe only for aqueous applications.

Take extra care when handling and storing the oxygen membrane 
module cap.

Do not allow the DO probe’s sensing area (cap reservoir) to dry out.

Perform the calibration procedure at the beginning of each day for 
maximum performance. Recalibrate the DO probe every two hours 
for maximum accuracy.

The DO probe can be left out of solution for up to 24 hours without 
damage. If the DO probe is not being used for more than 24 hours, 
prevent electrolyte evaporation by storing it in the storage/calibration 
chamber. Refer to the meter instruction manual for more information on 
probe storage. 

Probe Polarization 

Note: Prior to 
operation, connect the 
probe to the meter. 
Allow the probe to 
stabilize for 30-50 
minutes before 
attempting 
a calibration.

Hach dissolved oxygen probes are continuously polarized when they 
are connected to the instrument. A steady reading will not be seen for 
30–50 minutes when the probe is new or when it has been unplugged 
for more than an hour. Interrupted connections of less than one hour 
will require 5–25 minutes before stabilizing. Refer to the sens

ion

6 

Dissolved Oxygen Meter Instruction Manual for additional information. 

Probe Calibration

Detailed instructions for calibrating the Dissolved Oxygen Probe using 
a Water-saturated Air or Winkler Titration method are presented in the 
sens

ion

6 Dissolved Oxygen Meter Instruction Manual. Perform a 

calibration procedure at the beginning of each day for best performance. 
Obtain maximum accuracy by recalibrating the probe every two hours.

Содержание sension 6

Страница 1: ...sing the sension6 Dissolved Oxygen Meter Specifications Dissolved Oxygen Probe 1 meter cable Dissolved Oxygen Probe 3 meter cable Dissolved Oxygen Probe 15 meter cable Warranty 1 year 1 year 1 year Co...

Страница 2: ...sension6 Dissolved Oxygen Meter Instruction Manual for additional information Using the Weight Assembly The Weight Assembly is included with the probe equipped with the 15 m cable Install the weight...

Страница 3: ...or more information on probe storage Probe Polarization Note Prior to operation connect the probe to the meter Allow the probe to stabilize for 30 50 minutes before attempting a calibration Hach disso...

Страница 4: ...on around the probe tip by pulling on the cable to move the probe up and down approximately 3 6 inches 4 When the reading on the meter stabilizes record the value or store in meter memory Storage To s...

Страница 5: ...rmance of the probe seems to have degraded over time BOD Measurement Adapt the Model 51970 Dissolved Oxygen Probe for BOD measurements using the BOD Accessary Kit Cat No 51971 00 The kit includes a fu...

Страница 6: ...irrer 230 V without plug 45300 02 Filling Solution 50 mL 27591 26 Replacement Membranes 2 pkg 51973 00 Weight Assembly for the DO Probe 51969 00 Table 1 Elevation Barometric Pressure Elevation in feet...

Страница 7: ......

Страница 8: ...Hach Company 1999 All rights reserved Printed in the U S A dd dk 2 99 1ed...

Отзывы: