2

DISSOLVED OXYGEN PROBE, 

continued

Assembly Instructions

1. Hold the membrane module cap in a vertical position.

2. Fill the module cap about 

2

/

3

 of the way full with Dissolved Oxygen 

Electrolyte Filling Solution, Hach Cat. No. 27591-26.

3. Hold the DO probe in a vertical position with the tip down and 

gently screw the module cap onto the tip. Electrolyte should leak 
out of the threads. 

Note: If electrolyte does not leak out of the threads, air may remain inside the 

module cap. To ensure accurate results, repeat this procedure using more 
filling solution. 

Attach the DO probe cable connector to the input connector at the top of 
the meter. Refer to the sens

ion

6 Dissolved Oxygen Meter Instruction 

Manual for additional information. 

Using the Weight Assembly

The Weight Assembly is included with the probe equipped with the 
15 m cable. Install the weight assembly as follows:

1. Remove the cap from the body of the weight.

2. Insert the probe into the body. 

3. Slide the cap over the cable and thread the cap onto the body. 

The weight will help to ensure that the probe travels straight down after 
being placed in the water and also protects the membrane from damage. 

Zeroing the Probe

Chemically zeroing the probe is only necessary when measuring 
dissolved oxygen at a level less than 1 ppm (or 10% saturation). When 
measuring in this range, zero the probe prior to calibration. It is only 
necessary to zero a probe once. 

Refer to the sens

ion

6 Dissolved Oxygen Meter Instruction Manual for 

specific instructions. 

Содержание sension 6

Страница 1: ...sing the sension6 Dissolved Oxygen Meter Specifications Dissolved Oxygen Probe 1 meter cable Dissolved Oxygen Probe 3 meter cable Dissolved Oxygen Probe 15 meter cable Warranty 1 year 1 year 1 year Co...

Страница 2: ...sension6 Dissolved Oxygen Meter Instruction Manual for additional information Using the Weight Assembly The Weight Assembly is included with the probe equipped with the 15 m cable Install the weight...

Страница 3: ...or more information on probe storage Probe Polarization Note Prior to operation connect the probe to the meter Allow the probe to stabilize for 30 50 minutes before attempting a calibration Hach disso...

Страница 4: ...on around the probe tip by pulling on the cable to move the probe up and down approximately 3 6 inches 4 When the reading on the meter stabilizes record the value or store in meter memory Storage To s...

Страница 5: ...rmance of the probe seems to have degraded over time BOD Measurement Adapt the Model 51970 Dissolved Oxygen Probe for BOD measurements using the BOD Accessary Kit Cat No 51971 00 The kit includes a fu...

Страница 6: ...irrer 230 V without plug 45300 02 Filling Solution 50 mL 27591 26 Replacement Membranes 2 pkg 51973 00 Weight Assembly for the DO Probe 51969 00 Table 1 Elevation Barometric Pressure Elevation in feet...

Страница 7: ......

Страница 8: ...Hach Company 1999 All rights reserved Printed in the U S A dd dk 2 99 1ed...

Отзывы: