DISSOLVED OXYGEN PROBE, 

continued

Maintenance

Membrane cap replacement and refilling are required at scheduled 
intervals or if the probe does not function properly, e.g. in the case of a 
damaged membrane or contaminated DO probe. If the membrane is not 
damaged, the recommended time interval for replacing the Electrolyte 
Filling Solution is 1-2 months.

Note: Hach supplies a 
DO probe service kit 
that includes one 
50-mL bottle of 
Dissolved Oxygen 
Electrolyte Filling 
Solution, two 
Membrane Module 
Caps and a 
Polishing cloth.

Prior to replacing a membrane cap, rub the anode (the outer, metallic 
stem of the probe that is visible when the membrane cap is removed) 
with the polishing cloth supplied with the probe. The polishing cloth 
will remove deposits that may decrease the performance of the probe. 
Polish the anode whenever the membrane cap is replaced or between 
membrane cap replacements if the performance of the probe seems to 
have degraded over time. 

BOD Measurement

Adapt the Model 51970 Dissolved Oxygen Probe for BOD 
measurements using the BOD Accessary Kit (Cat. No. 51971-00). The 
kit includes a funnel and spacer that fit the Model 51970 DO Probe and 
a standard 300 mL BOD bottle. The funnel has a stir bar incorporated 
into the end which, when used with a stir stand, creates the proper 
stirring motion required for BOD measurement. The spacer positions 
the DO probe properly in the funnel. The funnel allows for sample 
displacement with insertion of the DO probe and limits exposure to 
atmospheric oxygen. Consult the sens

ion

6 Dissolved Oxygen Meter 

Instruction Manual for more information. 

Barometric Pressure Correction

The values presented in Table 1 are based on the assumption that the 
barometric pressure at sea level is 760 mm Hg. Use Table 1 to 
determine the barometric pressure at specific elevations or obtain the 
current barometric pressure from a local weather service. See the 
sens

ion

6 Dissolved Oxygen Meter Instruction Manual for specific 

calibration instructions when using Table 1

Содержание sension 6

Страница 1: ...sing the sension6 Dissolved Oxygen Meter Specifications Dissolved Oxygen Probe 1 meter cable Dissolved Oxygen Probe 3 meter cable Dissolved Oxygen Probe 15 meter cable Warranty 1 year 1 year 1 year Co...

Страница 2: ...sension6 Dissolved Oxygen Meter Instruction Manual for additional information Using the Weight Assembly The Weight Assembly is included with the probe equipped with the 15 m cable Install the weight...

Страница 3: ...or more information on probe storage Probe Polarization Note Prior to operation connect the probe to the meter Allow the probe to stabilize for 30 50 minutes before attempting a calibration Hach disso...

Страница 4: ...on around the probe tip by pulling on the cable to move the probe up and down approximately 3 6 inches 4 When the reading on the meter stabilizes record the value or store in meter memory Storage To s...

Страница 5: ...rmance of the probe seems to have degraded over time BOD Measurement Adapt the Model 51970 Dissolved Oxygen Probe for BOD measurements using the BOD Accessary Kit Cat No 51971 00 The kit includes a fu...

Страница 6: ...irrer 230 V without plug 45300 02 Filling Solution 50 mL 27591 26 Replacement Membranes 2 pkg 51973 00 Weight Assembly for the DO Probe 51969 00 Table 1 Elevation Barometric Pressure Elevation in feet...

Страница 7: ......

Страница 8: ...Hach Company 1999 All rights reserved Printed in the U S A dd dk 2 99 1ed...

Отзывы: